150 . Clarus 590 GC User’s Guide
NPD Overview
WARNING
AVERTISSEMENT
To avoid injury, Do Not turn on the hydrogen gas
unless a column is connected to the NPD and all
fitting connections have been leak tested
.
Nitrogen Phosphorus Detectors (NPD) use hydrogen gas.
Before you turn on the hydrogen gas to the NPD, ensure
that a column is connected to the NPD fitting inside the
oven. Turning on the hydrogen gas without first connecting
a column to the NPD could create the possibility of an
explosion in the oven.
Pour éviter les blessures, Ne pas allumer le gaz
hydrogène, sauf si une colonne est connectée à la
NPD et toutes les connexions de montage ont été
testées.
Les détecteurs de phosphore d'azote (NPD) utilisent de
l'hydrogène gazeux. Avant d'allumer l'hydrogène gaz au
NPD, assurez-vous qu'une colonne est connectée au raccord
NPD à l'intérieur du four. L'allumage de l'hydrogène gazeux
sans avoir préalablement relié une colonne au NPD pourrait
créer une explosion dans le four.
WARNING
AVERTISSEMENT
Turn off the hydrogen gas before disconnecting a
column from the NPD.
When two NPDs are installed, but only one NPD has a
column connected to it, plug the inlet of the NPD without a
column with a 1/8-inch stainless steel plug (P/N N9300061)
to prevent hydrogen from back-diffusing into the oven.
Éteignez le gaz hydrogène avant de débrancher une
colonne du NPD.
Lorsque deux NPD sont installés, mais un seul NPD a une
colonne qui lui est connectée, branchez l'entrée du NPD
sans colonne avec une prise en acier inoxydable de 1/8
pouces (P/N N9300061) pour empêcher l'hydrogène de
diffuser de nouveau dans le four.
NOTE: The detector gas modules have been set to OFF at the factory.
The nitrogen phosphorus detector (NPD) is a highly specific thermionic detector for organically bound
nitrogen and phosphorus.
NOTE: When using capillary columns with the NPD, Hydrogen is NOT a recommended carrier gas. The additive
effect of the hydrogen carrier gas and the hydrogen detector gas affects the sensitivity and selectivity
because it produces an overall higher hydrogen gas flow. If you are temperature programming, column
flow changes may produce a negative baseline drift.
Содержание CLARUS 590 GC
Страница 1: ...CLARUS 590 GC User s Guide GAS CHROMATOGRAPHY ...
Страница 9: ...About This Manual 9 Flame Ionization Detector FID 270 TCD 273 Autosampler 277 Index 279 ...
Страница 10: ...10 Clarus 590 GC User s Guide ...
Страница 11: ...1 Introduction ...
Страница 12: ...12 Clarus 590 GC User s Guide ...
Страница 21: ...2 Touch Screen Navigation ...
Страница 22: ...22 Clarus 590 GC User s Guide ...
Страница 48: ...48 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Страница 50: ...50 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Страница 59: ...3 Using the Active Method ...
Страница 60: ...60 Clarus 590 GC User s Guide ...
Страница 109: ...4 Setting Up the Detectors ...
Страница 110: ...110 Clarus 590 GC User s Guide ...
Страница 197: ...5 Using the Method Editor ...
Страница 198: ...198 Clarus 590 GC User s Guide ...
Страница 207: ...6 Using the Tools Menu ...
Страница 208: ...208 Clarus 590 GC User s Guide ...
Страница 224: ...224 Clarus 590 GC User s Guide ECD PID NPD FPD ...
Страница 225: ...Detectors 225 Output Configured ...
Страница 262: ...262 Clarus 590 GC User s Guide ...
Страница 263: ...7 Setting up a Typical Analysis ...
Страница 264: ...264 Clarus 590 GC User s Guide ...
Страница 279: ...Index ...
Страница 280: ...280 Clarus 590 GC User s Guide ...