Utilities . 257
Column Flow Calc Main Screen
Length
This field is normally entered in meters (1 to 1000 in 1m increments). The calculator should also support
techniques such as PreVent and heart-cutting etc. and so shorter columns/tubes should be supported. If
the value scrolls below 1m, then the units should change to mm (1 to 1000 in 1mm increments).
The default value will be 30m.
For the calculation this value should be converted to cm.
Diameter
Column diameter should be entered in µm (1 to 1000µm in 1µm increments).
The default will be 320µm.
Value converted to cm for calculations.
Temp
Normal GC oven temperature range in °C (-99 to 450°C in 1°C increments).
The default will be 50°C.
Value converted to °K for calculations.
Carrier Gas Name
The name of the carrier gas is selected in order to apply the correct viscosity coefficients and suggested
column carrier gas velocities as shown in the Table.
Screen Name
Gas
ref
η
,Pa.s
x
Suggested
Velocity, cm/s
He
Helium
0.000018662
0.646
20 to 40
H2
Hydrogen
0.000008399
0.725
30 to 55
N2
Nitrogen
0.000016736
0.680
8 to 16
The default will be “He”.
Outlet Press
The column outlet pressure is set to Ambient or Vacuum (0) or to a user entered Other value. Units will
be in absolute units (psi abs or kPa abs) and will be set according to the units chosen in the Configuration
screens for the rest of the user interaction. The range for this field will be 0 to 100 psi abs in 0.1psi
increments or 0 to 689 kPa abs in 1kPa increments.
The default will be Ambient (14.7psi abs or 101kPa abs).
Suggested Velocity
A display-only field that shows the recommended carrier gas velocity range from the selected carrier gas
name.
Inlet Press
This is a selectable field (this is the default selection) and, if selected, the Flow and Velocity/t0 fields are
deselected and are display only.
The carrier gas input pressure is in units of psig (1 to 100psig in 0.1psi increments) or kPa (1to 689kPa in
1kPa increments).
Содержание CLARUS 590 GC
Страница 1: ...CLARUS 590 GC User s Guide GAS CHROMATOGRAPHY ...
Страница 9: ...About This Manual 9 Flame Ionization Detector FID 270 TCD 273 Autosampler 277 Index 279 ...
Страница 10: ...10 Clarus 590 GC User s Guide ...
Страница 11: ...1 Introduction ...
Страница 12: ...12 Clarus 590 GC User s Guide ...
Страница 21: ...2 Touch Screen Navigation ...
Страница 22: ...22 Clarus 590 GC User s Guide ...
Страница 48: ...48 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Страница 50: ...50 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Страница 59: ...3 Using the Active Method ...
Страница 60: ...60 Clarus 590 GC User s Guide ...
Страница 109: ...4 Setting Up the Detectors ...
Страница 110: ...110 Clarus 590 GC User s Guide ...
Страница 197: ...5 Using the Method Editor ...
Страница 198: ...198 Clarus 590 GC User s Guide ...
Страница 207: ...6 Using the Tools Menu ...
Страница 208: ...208 Clarus 590 GC User s Guide ...
Страница 224: ...224 Clarus 590 GC User s Guide ECD PID NPD FPD ...
Страница 225: ...Detectors 225 Output Configured ...
Страница 262: ...262 Clarus 590 GC User s Guide ...
Страница 263: ...7 Setting up a Typical Analysis ...
Страница 264: ...264 Clarus 590 GC User s Guide ...
Страница 279: ...Index ...
Страница 280: ...280 Clarus 590 GC User s Guide ...