126 . Clarus 590 GC User’s Guide
Using the FID in Corrosive Environments
If you are analyzing samples (such as chloroform, methylene chloride, etc.), that produce corrosive by-
products, we recommend installing the piece of vent tubing (P/N 02506490) supplied in the Clarus GC
Shipping Kit (P/N N6100348) and setting the FID temperature greater than 250° C to minimize
condensation from forming. Since these corrosive by-products attack the top of the FID collector assembly,
including the ignitor, we strongly recommend venting the FID into a vent hood.
NOTE: The flame will not ignite with this vent tubing installed. To ignite the flame, you must first remove the vent
tubing, ignite the flame, and then replace the vent tubing.
To install the vent tubing:
1.
Ignite the FID flame.
WARNING
AVERTISSEMENT
The FID outlet is hot! To prevent injury, never place
your hand over the FID outlet.
La sortie FID est chaude! Pour éviter toute blessure, ne
placez jamais votre main sur la prise FID.
2.
Lift-open the detector cover.
3.
Cut the piece of tubing (P/N 02506490) to a length of 4-inches and firmly insert one end of the
tubing into the top of the FID.
4.
Route the end of the tubing through the hole in the detector cover as you close the detector cover.
5.
Connect one end of a length of vent tubing to this piece of tubing and place the other end in a vent
hood.
Содержание CLARUS 590 GC
Страница 1: ...CLARUS 590 GC User s Guide GAS CHROMATOGRAPHY ...
Страница 9: ...About This Manual 9 Flame Ionization Detector FID 270 TCD 273 Autosampler 277 Index 279 ...
Страница 10: ...10 Clarus 590 GC User s Guide ...
Страница 11: ...1 Introduction ...
Страница 12: ...12 Clarus 590 GC User s Guide ...
Страница 21: ...2 Touch Screen Navigation ...
Страница 22: ...22 Clarus 590 GC User s Guide ...
Страница 48: ...48 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Страница 50: ...50 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Страница 59: ...3 Using the Active Method ...
Страница 60: ...60 Clarus 590 GC User s Guide ...
Страница 109: ...4 Setting Up the Detectors ...
Страница 110: ...110 Clarus 590 GC User s Guide ...
Страница 197: ...5 Using the Method Editor ...
Страница 198: ...198 Clarus 590 GC User s Guide ...
Страница 207: ...6 Using the Tools Menu ...
Страница 208: ...208 Clarus 590 GC User s Guide ...
Страница 224: ...224 Clarus 590 GC User s Guide ECD PID NPD FPD ...
Страница 225: ...Detectors 225 Output Configured ...
Страница 262: ...262 Clarus 590 GC User s Guide ...
Страница 263: ...7 Setting up a Typical Analysis ...
Страница 264: ...264 Clarus 590 GC User s Guide ...
Страница 279: ...Index ...
Страница 280: ...280 Clarus 590 GC User s Guide ...