138 . Clarus 590 GC User’s Guide
CAUTION
ATTENTION
The TCD flow module for reference gas is calibrated with
helium gas at the factory. You must recalibrate the TCD flow
module if you are using a reference gas other than helium.
Le module de débit TCD pour gaz de référence est étalonné
avec de l'hélium gaz en usine. Vous devez recalibrer le module
de flux TCD si vous utilisez un gaz de référence autre que
l'hélium.
For optimum performance, apply a minimum flow of 5 mL/min to each of the two detector channels.
Set the Makeup and Reference Gas in the PPC Version
With PPC pneumatics the reference gas is set from the Clarus GC touch screen. Also, PPC pneumatics
provide the user with TCD filament protection by monitoring the carrier gas leak tests or monitoring the gas
through the alarm feature and shutting down the TCD current if a gas leak or no gas at all is detected.
The PPC modules are calibrated at the factory; therefore, to set the flows for an analysis, simply follow this
procedure.
TCD Active Method
TCD Configuration Screen
1.
To set the make-up and reference gases touch the
button in TCD tab
screen then touch
Configuration
from the drop down menu.
This will bring you to the TCD Configuration screen page (see screens above).
2.
In the TCD Configuration screen set the make-up gas and reference gas line
pressures to a range of
60 - 90
psig by using the plus minus buttons.
3.
As shown in the example above use the plus minus buttons set the filter to
200
and the Output to
5
.When all the settings are appropriate for your analysis
touch
OK
. The TCD tab screen will be now visible and the changes you made in the
configuration screen will be shown.
The default setting for the filter is 200 milliseconds. If you want to improve the signal-to-noise, change the
setting to 800 milliseconds; however, the noise improves but the peak shapes broaden. If you want a
faster detector response, change the setting to 50 milliseconds; however, the noise level may increase.
Содержание CLARUS 590 GC
Страница 1: ...CLARUS 590 GC User s Guide GAS CHROMATOGRAPHY ...
Страница 9: ...About This Manual 9 Flame Ionization Detector FID 270 TCD 273 Autosampler 277 Index 279 ...
Страница 10: ...10 Clarus 590 GC User s Guide ...
Страница 11: ...1 Introduction ...
Страница 12: ...12 Clarus 590 GC User s Guide ...
Страница 21: ...2 Touch Screen Navigation ...
Страница 22: ...22 Clarus 590 GC User s Guide ...
Страница 48: ...48 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Страница 50: ...50 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Страница 59: ...3 Using the Active Method ...
Страница 60: ...60 Clarus 590 GC User s Guide ...
Страница 109: ...4 Setting Up the Detectors ...
Страница 110: ...110 Clarus 590 GC User s Guide ...
Страница 197: ...5 Using the Method Editor ...
Страница 198: ...198 Clarus 590 GC User s Guide ...
Страница 207: ...6 Using the Tools Menu ...
Страница 208: ...208 Clarus 590 GC User s Guide ...
Страница 224: ...224 Clarus 590 GC User s Guide ECD PID NPD FPD ...
Страница 225: ...Detectors 225 Output Configured ...
Страница 262: ...262 Clarus 590 GC User s Guide ...
Страница 263: ...7 Setting up a Typical Analysis ...
Страница 264: ...264 Clarus 590 GC User s Guide ...
Страница 279: ...Index ...
Страница 280: ...280 Clarus 590 GC User s Guide ...