106 . Clarus 590 GC User’s Guide
Cleaning the 5-
µ
L and 50-
µ
L Syringe Plungers
The 5-
µ
L and 50-
µ
L syringe plungers should be cleaned regularly, since insolubles can build up and cause
friction.
To clean the syringe plunger:
1.
Remove the syringe using the procedure in the preceding section.
2.
Remove the plunger from the syringe barrel.
3.
Wipe the plunger with a tissue soaked in an appropriate solvent. Replace the plunger.
4.
Pull and expel the same solvent through the barrel several times.
5.
Replace the syringe using the procedure in the preceding section.
NOTE: Only syringes distributed by PerkinElmer should be used with the Clarus GC. Plungers are not
interchangeable from syringe to syringe.
Configuration Button
From the
Tasks
tab touch the
Config
button.
The
Config
button provides a quick link to the configuration settings.
Auto Resume
Check the
Auto Resume
checkbox to allow the autosampler program to automatically resume where it left
off should a power failure occur.
NOTE: Auto Resume should not be used if you have an FID or FPD configured with PPC, since the flame will have
been extinguished during a power failure.
Running Viscous Samples
The Clarus GC allows you to vary the length of time the plunger pauses at the top-of-stroke when acquiring
samples. In addition, the plunger speed also slows down during injection. The more viscous the sample,
the higher you should set this value. To optimize reproducibility you should experiment with different
values.
Содержание CLARUS 590 GC
Страница 1: ...CLARUS 590 GC User s Guide GAS CHROMATOGRAPHY ...
Страница 9: ...About This Manual 9 Flame Ionization Detector FID 270 TCD 273 Autosampler 277 Index 279 ...
Страница 10: ...10 Clarus 590 GC User s Guide ...
Страница 11: ...1 Introduction ...
Страница 12: ...12 Clarus 590 GC User s Guide ...
Страница 21: ...2 Touch Screen Navigation ...
Страница 22: ...22 Clarus 590 GC User s Guide ...
Страница 48: ...48 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Страница 50: ...50 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Страница 59: ...3 Using the Active Method ...
Страница 60: ...60 Clarus 590 GC User s Guide ...
Страница 109: ...4 Setting Up the Detectors ...
Страница 110: ...110 Clarus 590 GC User s Guide ...
Страница 197: ...5 Using the Method Editor ...
Страница 198: ...198 Clarus 590 GC User s Guide ...
Страница 207: ...6 Using the Tools Menu ...
Страница 208: ...208 Clarus 590 GC User s Guide ...
Страница 224: ...224 Clarus 590 GC User s Guide ECD PID NPD FPD ...
Страница 225: ...Detectors 225 Output Configured ...
Страница 262: ...262 Clarus 590 GC User s Guide ...
Страница 263: ...7 Setting up a Typical Analysis ...
Страница 264: ...264 Clarus 590 GC User s Guide ...
Страница 279: ...Index ...
Страница 280: ...280 Clarus 590 GC User s Guide ...