82 . Clarus 590 GC User’s Guide
Now you can only pressure program your capillary injector. If you wish to program the Capillary
control for more than just pressure skip to step 9.
4.
Under
Flow Offset
, select either the
Auto
or
Fixed
.
The
Auto
mode is the preferred mode of operation. If you want to shorten the overall run time at
the expense of some area precision, then use the
Fixed
mode.
We recommend that if you are operating with a split vent flow of 50 mL/min or less, select the
Auto
mode.
Before each injection in the
Auto
mode, the system briefly closes the split vent and measures the
total flow. With the split vent off this flow is the column flow plus the septum purge flow. The
septum purge flow is maintained due to a mechanical restrictor and the split vent flow and split
ratio are maintained for each injection.
In the
Fixed
mode, you enter an offset value which is equal to the septum purge flow plus the
column flow at the time of injection. Since the system controls the total flow (equal to the split
vent plus septum purge plus column flow), the septum purge is fixed, and the column flow is
controlled, the difference between these two flows and the offset entered will go out the vent flow.
This will be consistent for each run.
5.
If
Fixed
is selected use the up and down arrow or keypad
buttons to enter an offset
value in mL/min.
6.
Press
OK
.
Содержание CLARUS 590 GC
Страница 1: ...CLARUS 590 GC User s Guide GAS CHROMATOGRAPHY ...
Страница 9: ...About This Manual 9 Flame Ionization Detector FID 270 TCD 273 Autosampler 277 Index 279 ...
Страница 10: ...10 Clarus 590 GC User s Guide ...
Страница 11: ...1 Introduction ...
Страница 12: ...12 Clarus 590 GC User s Guide ...
Страница 21: ...2 Touch Screen Navigation ...
Страница 22: ...22 Clarus 590 GC User s Guide ...
Страница 48: ...48 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Страница 50: ...50 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Страница 59: ...3 Using the Active Method ...
Страница 60: ...60 Clarus 590 GC User s Guide ...
Страница 109: ...4 Setting Up the Detectors ...
Страница 110: ...110 Clarus 590 GC User s Guide ...
Страница 197: ...5 Using the Method Editor ...
Страница 198: ...198 Clarus 590 GC User s Guide ...
Страница 207: ...6 Using the Tools Menu ...
Страница 208: ...208 Clarus 590 GC User s Guide ...
Страница 224: ...224 Clarus 590 GC User s Guide ECD PID NPD FPD ...
Страница 225: ...Detectors 225 Output Configured ...
Страница 262: ...262 Clarus 590 GC User s Guide ...
Страница 263: ...7 Setting up a Typical Analysis ...
Страница 264: ...264 Clarus 590 GC User s Guide ...
Страница 279: ...Index ...
Страница 280: ...280 Clarus 590 GC User s Guide ...