ECD Overview . 131
NOTE: Because an ECD is extremely sensitive, contamination from any part of the system can add to the
background and noise. To help assure a clean system the injector and detector should be baked-out and
both packed and capillary columns thoroughly conditioned before use. When conditioning columns for use
with an ECD, DO NOT attach the column to the ECD. Clean tubing and pure filtered gases should be used.
Whenever a column is disconnected from an ECD, the ECD should first be cooled to avoid oxidation of the
Nickel foil.
Vent the ECD Outlet
NOTE: If venting the ECD effluent is required, attach an additional piece of tubing to the tubing that is connected
to the ECD outlet. This does not affect the performance of the ECD.
1.
Open the detector cover (see the following figure) to expose the top of the ECD.
2.
Remove the two screws that secure the insulating cover to the Clarus GC.
3.
Remove the insulating cover and detector cap.
This exposes the ECD outlet (see the following figure).
4.
Attach the outer metal sleeve of the vent tubing (P/N N6100161) to the ECD Outlet.
5.
Press down and turn the metal sleeve on the end of the vent tubing to
secure it to the ECD.
6.
Thread the vent tubing through the detector cap and insulating cover as you position them over
the ECD.
Содержание CLARUS 590 GC
Страница 1: ...CLARUS 590 GC User s Guide GAS CHROMATOGRAPHY ...
Страница 9: ...About This Manual 9 Flame Ionization Detector FID 270 TCD 273 Autosampler 277 Index 279 ...
Страница 10: ...10 Clarus 590 GC User s Guide ...
Страница 11: ...1 Introduction ...
Страница 12: ...12 Clarus 590 GC User s Guide ...
Страница 21: ...2 Touch Screen Navigation ...
Страница 22: ...22 Clarus 590 GC User s Guide ...
Страница 48: ...48 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Страница 50: ...50 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Страница 59: ...3 Using the Active Method ...
Страница 60: ...60 Clarus 590 GC User s Guide ...
Страница 109: ...4 Setting Up the Detectors ...
Страница 110: ...110 Clarus 590 GC User s Guide ...
Страница 197: ...5 Using the Method Editor ...
Страница 198: ...198 Clarus 590 GC User s Guide ...
Страница 207: ...6 Using the Tools Menu ...
Страница 208: ...208 Clarus 590 GC User s Guide ...
Страница 224: ...224 Clarus 590 GC User s Guide ECD PID NPD FPD ...
Страница 225: ...Detectors 225 Output Configured ...
Страница 262: ...262 Clarus 590 GC User s Guide ...
Страница 263: ...7 Setting up a Typical Analysis ...
Страница 264: ...264 Clarus 590 GC User s Guide ...
Страница 279: ...Index ...
Страница 280: ...280 Clarus 590 GC User s Guide ...