180 . Clarus 590 GC User’s Guide
3.
To turn on the photomultiplier tube touch the
button in FPD tab screen then touch
Configuration
from the drop down menu . This will bring you to the FPD Configuration screen
page.
4.
In the PMT Voltage field set the percentage (see note below) by using the plus minus buttons. This
will turn on the photomultiplier tube.
NOTE: The voltage to the photomultiplier tube has been set in the factory to approximately 70%. You can change
this setting from the Clarus GC touchscreen. You may want to change this setting as the photomultiplier
tube ages and the signal diminishes. The signal-to-noise ratio may vary with each photomultiplier tube;
therefore, increasing the voltage may not increase the signal-to-noise ratio.
After installing a replacement photomultiplier tube, the noise should be about 2% when the Clarus GC
integrator output is set to attenuation 8 (x8) and the linearizer off. See the
Clarus 590 GC Hardware Guide
"Maintenance” chapter, for this procedure.
NOTE: The photomultiplier tube can be damaged if exposed to light while voltage is ON. Never remove the
detector cap or Photomultiplier tube WITH the voltage on.
Ignite the Flame
1.
Open the electronics cover to access the ignitor assembly.
2.
From the system status screen touch the
FPD
icon.
3.
The FPD tab page is now visible.
Содержание CLARUS 590 GC
Страница 1: ...CLARUS 590 GC User s Guide GAS CHROMATOGRAPHY ...
Страница 9: ...About This Manual 9 Flame Ionization Detector FID 270 TCD 273 Autosampler 277 Index 279 ...
Страница 10: ...10 Clarus 590 GC User s Guide ...
Страница 11: ...1 Introduction ...
Страница 12: ...12 Clarus 590 GC User s Guide ...
Страница 21: ...2 Touch Screen Navigation ...
Страница 22: ...22 Clarus 590 GC User s Guide ...
Страница 48: ...48 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Страница 50: ...50 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Страница 59: ...3 Using the Active Method ...
Страница 60: ...60 Clarus 590 GC User s Guide ...
Страница 109: ...4 Setting Up the Detectors ...
Страница 110: ...110 Clarus 590 GC User s Guide ...
Страница 197: ...5 Using the Method Editor ...
Страница 198: ...198 Clarus 590 GC User s Guide ...
Страница 207: ...6 Using the Tools Menu ...
Страница 208: ...208 Clarus 590 GC User s Guide ...
Страница 224: ...224 Clarus 590 GC User s Guide ECD PID NPD FPD ...
Страница 225: ...Detectors 225 Output Configured ...
Страница 262: ...262 Clarus 590 GC User s Guide ...
Страница 263: ...7 Setting up a Typical Analysis ...
Страница 264: ...264 Clarus 590 GC User s Guide ...
Страница 279: ...Index ...
Страница 280: ...280 Clarus 590 GC User s Guide ...