146 . Clarus 590 GC User’s Guide
Set the Make-Up Gas in the PPC Version
CAUTION
ATTENTION
The PID flow module for reference gas is calibrated with helium
gas at the factory. You must recalibrate the PID flow module if
you are using a reference gas other than helium.
Le module de débit PID pour le gaz de référence est étalonné avec
de l'hélium gaz en usine. Vous devez recalibrer le module de flux
PID si vous utilisez un gaz de référence autre que l'hélium.
1.
With PPC pneumatics the make-up gas is set from the Clarus GC touch screen.
The PPC modules are calibrated at the factory: therefore, to set the flows for an analysis, simply
follow this procedure.
2.
To set the make-up and reference gases touch the
button in PID tab screen
then touch
Configuration
from the drop down menu. This will bring you to the
PID Configuration screen page (see screens above).
3.
In the PID Configuration screen set the make-up gas and reference gas line
pressures to 60 - 90 psig by using the up and down arrow or keypad
buttons.
4.
As shown in the example above use the up and down arrow or keypad
buttons to set the filter to 200 and the output to 5.When all the settings are
appropriate for your analysis touch
OK
. The TCD tab screen will be now visible and the changes
you made in the configuration screen will be shown.
The default setting for the filter is 200 milliseconds. If you want to improve the signal-to-noise,
change the setting to 800 milliseconds; however, the noise improves but the peak shapes broaden.
If you want a faster detector response, change the setting to 50 milliseconds; however, the noise
level may increase.
Adjust the Make-Up Gas in the Manual Pneumatics Version
To adjust the make-up gas in the manual pneumatics version use the on/off needle valve mounted on the
pneumatics control panel.
NOTE: Do not overtighten the inner flow adjustment screw in the on/off needle valve. Overtightening damages
the mechanical device.
The manual pneumatics make-up gas control valve is shown below.
The make-up gas control valve consists of an outer ring (used to turn the make-up gas completely off or
on), a protective cap, and an inner needle valve (used to adjust the flow of make-up gas). Removing the
Содержание CLARUS 590 GC
Страница 1: ...CLARUS 590 GC User s Guide GAS CHROMATOGRAPHY ...
Страница 9: ...About This Manual 9 Flame Ionization Detector FID 270 TCD 273 Autosampler 277 Index 279 ...
Страница 10: ...10 Clarus 590 GC User s Guide ...
Страница 11: ...1 Introduction ...
Страница 12: ...12 Clarus 590 GC User s Guide ...
Страница 21: ...2 Touch Screen Navigation ...
Страница 22: ...22 Clarus 590 GC User s Guide ...
Страница 48: ...48 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Страница 50: ...50 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Страница 59: ...3 Using the Active Method ...
Страница 60: ...60 Clarus 590 GC User s Guide ...
Страница 109: ...4 Setting Up the Detectors ...
Страница 110: ...110 Clarus 590 GC User s Guide ...
Страница 197: ...5 Using the Method Editor ...
Страница 198: ...198 Clarus 590 GC User s Guide ...
Страница 207: ...6 Using the Tools Menu ...
Страница 208: ...208 Clarus 590 GC User s Guide ...
Страница 224: ...224 Clarus 590 GC User s Guide ECD PID NPD FPD ...
Страница 225: ...Detectors 225 Output Configured ...
Страница 262: ...262 Clarus 590 GC User s Guide ...
Страница 263: ...7 Setting up a Typical Analysis ...
Страница 264: ...264 Clarus 590 GC User s Guide ...
Страница 279: ...Index ...
Страница 280: ...280 Clarus 590 GC User s Guide ...