FID Overview . 117
The PPC version of the FIDW has
flameout
detection and an
auto-ignite
feature. Flameout detection
occurs 0.5 minutes into the PRE-RUN time and it measures the user-entered baseline threshold level in mV.
If the threshold level is exceeded, the flame is considered lit, otherwise the Clarus GC detects that the
flame is out and auto-ignite automatically lights the flame before the start of a run.
With the FIDW temperature set, the air and hydrogen flows set, and auto-ignition activated by setting a
positive mV value for the flame offset, the FIDW attempts auto-ignition at -0.5 min. before the system
becomes ready. At this Pre-run time the status bar flashes “
Flame Ignition”
as the air flow is shut off,
hydrogen purges the FIDW for about 10 seconds, and the ignitor energizes for ignition followed by a return
of the air flow to the detector to sustain the flame.
After the flame ignites, the FIDW output level is monitored to confirm ignition before producing a READY
signal. If ignition is not confirmed, the ignition process repeats up to two more times if needed. If ignition
is not achieved after three attempts, the flame gases are set to off and a dialog box appears displays the
following message:
“
Flame 1 or 2 failed to ignite.”
Touch the OK button. The status is made NOT READY and “
FID 1 or 2 flameout”
appears. If auto-
ignition is attempted with the FIDW gases set to zero flow, the message “
No ignition. Flame gases are
off”
appears on the screen. To proceed, go to the Configuration screen and reset the air and hydrogen to
their 450 and 30 mL/min flowrates.
To Auto-Ignite
The auto-ignition process only occurs before the start of each run just prior to the instrument becoming
READY. However, the manual flame ignition process can be initiated at any time by touching the
Ignite
button
.
1.
When auto-ignite begins the Clarus GC equilibrates.
2.
Once the two minutes of equilibration expire,
Pre-run
begins and the GC detects that the flame is
not lit and indicates this by alternately flashing
Flame
and the
method name
in the status bar.
3.
After determining that the flame is not lit, the GC switches to a not ready state and attempts to
automatically ignite the flame.
Flame Ignition 1
appears in the status bar.
4.
The air flow is shut-off and hydrogen is purged through the FID for
10 seconds.
5.
The ignitor energizes, the flame ignites (indicated by the POP), and air flow is returned to sustain
the flame. The FID output level is monitored to confirm that the flame is lit. If a lighted flame is
not confirmed, the Clarus GC repeats the ignition process up to two more times.
Manually Ignite the Flame
You can manually ignite the flame at any time by following this procedure. Before trying to ignite the
flame, make sure that you have sufficient combustion gases and they are turned on at the source, and
make sure that the flows are set as described earlier in this section.
Содержание CLARUS 590 GC
Страница 1: ...CLARUS 590 GC User s Guide GAS CHROMATOGRAPHY ...
Страница 9: ...About This Manual 9 Flame Ionization Detector FID 270 TCD 273 Autosampler 277 Index 279 ...
Страница 10: ...10 Clarus 590 GC User s Guide ...
Страница 11: ...1 Introduction ...
Страница 12: ...12 Clarus 590 GC User s Guide ...
Страница 21: ...2 Touch Screen Navigation ...
Страница 22: ...22 Clarus 590 GC User s Guide ...
Страница 48: ...48 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Страница 50: ...50 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Страница 59: ...3 Using the Active Method ...
Страница 60: ...60 Clarus 590 GC User s Guide ...
Страница 109: ...4 Setting Up the Detectors ...
Страница 110: ...110 Clarus 590 GC User s Guide ...
Страница 197: ...5 Using the Method Editor ...
Страница 198: ...198 Clarus 590 GC User s Guide ...
Страница 207: ...6 Using the Tools Menu ...
Страница 208: ...208 Clarus 590 GC User s Guide ...
Страница 224: ...224 Clarus 590 GC User s Guide ECD PID NPD FPD ...
Страница 225: ...Detectors 225 Output Configured ...
Страница 262: ...262 Clarus 590 GC User s Guide ...
Страница 263: ...7 Setting up a Typical Analysis ...
Страница 264: ...264 Clarus 590 GC User s Guide ...
Страница 279: ...Index ...
Страница 280: ...280 Clarus 590 GC User s Guide ...