ECD Overview . 133
The following screen appears.
The default setting is 200 milliseconds. If you want to improve the signal-to-noise, change the
setting to 800 milliseconds; however, the noise improves but the peak shapes broaden. If you
want a faster detector response, change the setting to 50 milliseconds; however, the noise level
may increase.
5.
In this Configuration screen use the up and down arrow or keypad
buttons to set the
Make-up
gas flow.
6.
Touch
OK
when finished.
Adjust the Make-Up Gas in the Manual Pneumatics Version
To adjust the make-up gas in the manual pneumatics version, use the on/off needle valve mounted on the
pneumatics control panel.
CAUTION
Do not overtighten the inner flow adjustment screw in the on/off
needle valve or the on/off regulator. Overtightening damages the
mechanical device.
Содержание CLARUS 590 GC
Страница 1: ...CLARUS 590 GC User s Guide GAS CHROMATOGRAPHY ...
Страница 9: ...About This Manual 9 Flame Ionization Detector FID 270 TCD 273 Autosampler 277 Index 279 ...
Страница 10: ...10 Clarus 590 GC User s Guide ...
Страница 11: ...1 Introduction ...
Страница 12: ...12 Clarus 590 GC User s Guide ...
Страница 21: ...2 Touch Screen Navigation ...
Страница 22: ...22 Clarus 590 GC User s Guide ...
Страница 48: ...48 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Страница 50: ...50 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Страница 59: ...3 Using the Active Method ...
Страница 60: ...60 Clarus 590 GC User s Guide ...
Страница 109: ...4 Setting Up the Detectors ...
Страница 110: ...110 Clarus 590 GC User s Guide ...
Страница 197: ...5 Using the Method Editor ...
Страница 198: ...198 Clarus 590 GC User s Guide ...
Страница 207: ...6 Using the Tools Menu ...
Страница 208: ...208 Clarus 590 GC User s Guide ...
Страница 224: ...224 Clarus 590 GC User s Guide ECD PID NPD FPD ...
Страница 225: ...Detectors 225 Output Configured ...
Страница 262: ...262 Clarus 590 GC User s Guide ...
Страница 263: ...7 Setting up a Typical Analysis ...
Страница 264: ...264 Clarus 590 GC User s Guide ...
Страница 279: ...Index ...
Страница 280: ...280 Clarus 590 GC User s Guide ...