108 . Clarus 590 GC User’s Guide
Wash vials
Program 1
splits between solvent wash vials #1 and #2. For example, if four washes are required, two
will be taken from vial #1, and two will be taken from vial #2.
Program 2
splits between solvent wash vials #3 and #4. For example, if five washes are required, three
will be taken from vial #3, and two will be taken from vial #4.
About Sample Vials
Use 2-mL vials with the autosampler (P/N N9301385, 144 per box). These are crimp-top vials. The vials
and crimp caps with septa (P/N 00092205, 1000 per box) are available through PerkinElmer’s Consumables
and Supplies catalog (in the U.S., call: 1-800-762-4000) or online at
www.perkinelmer.com
.
The minimum volume of sample required in the 2-mL vial is 0.35 mL or 350
µ
L. You should fill the vial
when running multiple injections per vial.
A mini-vial insert is available for laboratories with small sample volume requirements. This insert fits into
the standard 2-mL vial and requires a minimum volume of 5
µ
L of sample. These inserts are available
through the Consumables and Supplies catalog or visit us on-line at
www.perkinelmer.com
.
NOTE: When using mini-vials, it is not recommended to use pumps to acquire samples. This assures that the
sample will not be diluted by residual solvent in the syringe.
Only PerkinElmer vials should be used with the Clarus GC.
Adjust the crimper so that it does not deform the vial cap and septum, but crimps the vial cap tight enough
so that you cannot turn the cap. An improper adjusted crimper can cause the autosampler to malfunction.
Using a TCD w ith the Autosampler
When using a TCD at high sensitivity with a GSV or the autosampler, a minor baseline upset may occur at
the time of injection.
If the signal goes below zero, incorporate a timed event Autozero to bring the signal back to zero before
the peaks of interest elute.
Observe the initial chromatogram to determine the proper time for the timed event.
Available syringes
The following syringes are available for the Clarus GC:
Syringe
Part Number
5-µL (0.63 mm O.D.) Teflon-tipped plunger (Std)
N6101390
0.5-µL (0.63 mm O.D.)
N6101252
0.5-µL (0.47 mm O.D.)
N6101253
5-µL (0.63 mm O.D.)
N6101251
5-µL (0.47 mm O.D.)
N6101380
50-µL (0.63 mm O.D.)
N6101760
NOTE: A 0.47-mm needle is required for use with the POC injector and 530-
µ
m columns in the on-column injection
mode only. You can achieve better precision in the on-column injection mode when you use sample
volumes of 1.0
µ
L or greater.
Содержание CLARUS 590 GC
Страница 1: ...CLARUS 590 GC User s Guide GAS CHROMATOGRAPHY ...
Страница 9: ...About This Manual 9 Flame Ionization Detector FID 270 TCD 273 Autosampler 277 Index 279 ...
Страница 10: ...10 Clarus 590 GC User s Guide ...
Страница 11: ...1 Introduction ...
Страница 12: ...12 Clarus 590 GC User s Guide ...
Страница 21: ...2 Touch Screen Navigation ...
Страница 22: ...22 Clarus 590 GC User s Guide ...
Страница 48: ...48 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Страница 50: ...50 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Страница 59: ...3 Using the Active Method ...
Страница 60: ...60 Clarus 590 GC User s Guide ...
Страница 109: ...4 Setting Up the Detectors ...
Страница 110: ...110 Clarus 590 GC User s Guide ...
Страница 197: ...5 Using the Method Editor ...
Страница 198: ...198 Clarus 590 GC User s Guide ...
Страница 207: ...6 Using the Tools Menu ...
Страница 208: ...208 Clarus 590 GC User s Guide ...
Страница 224: ...224 Clarus 590 GC User s Guide ECD PID NPD FPD ...
Страница 225: ...Detectors 225 Output Configured ...
Страница 262: ...262 Clarus 590 GC User s Guide ...
Страница 263: ...7 Setting up a Typical Analysis ...
Страница 264: ...264 Clarus 590 GC User s Guide ...
Страница 279: ...Index ...
Страница 280: ...280 Clarus 590 GC User s Guide ...