FPD Overview . 167
3.
Use the up and down arrow or keypad
buttons
to e nte r a se tpoint of a t le a st 250 C
(this is the minimum entry for a FPD) in the temp field
, or enter a temperature that is
50
°
C higher than the highest temperature of your oven program.
Set the Hydrogen Flow and Air Flow for PPC
The PPC pneumatics consists of PPC modules where the hydrogen and air gases are set from the Clarus GC
touch screen.
The PPC modules are calibrated at the factory, therefore to set the flows for an analysis, simply follow this
procedure.
1.
Turn on the air at the source and adjust the pressure to 70 - 90 psig.
2.
Turn on the hydrogen at the source and adjust the pressure to 60 - 90 psig.
3.
From the System Status screen touch the FPD icon.
The FPD tab page is now visible.
4.
To change the gas flow values touch the
button in FPD tab screen then touch
Configuration
from the drop down menu.
Содержание CLARUS 590 GC
Страница 1: ...CLARUS 590 GC User s Guide GAS CHROMATOGRAPHY ...
Страница 9: ...About This Manual 9 Flame Ionization Detector FID 270 TCD 273 Autosampler 277 Index 279 ...
Страница 10: ...10 Clarus 590 GC User s Guide ...
Страница 11: ...1 Introduction ...
Страница 12: ...12 Clarus 590 GC User s Guide ...
Страница 21: ...2 Touch Screen Navigation ...
Страница 22: ...22 Clarus 590 GC User s Guide ...
Страница 48: ...48 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Страница 50: ...50 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Страница 59: ...3 Using the Active Method ...
Страница 60: ...60 Clarus 590 GC User s Guide ...
Страница 109: ...4 Setting Up the Detectors ...
Страница 110: ...110 Clarus 590 GC User s Guide ...
Страница 197: ...5 Using the Method Editor ...
Страница 198: ...198 Clarus 590 GC User s Guide ...
Страница 207: ...6 Using the Tools Menu ...
Страница 208: ...208 Clarus 590 GC User s Guide ...
Страница 224: ...224 Clarus 590 GC User s Guide ECD PID NPD FPD ...
Страница 225: ...Detectors 225 Output Configured ...
Страница 262: ...262 Clarus 590 GC User s Guide ...
Страница 263: ...7 Setting up a Typical Analysis ...
Страница 264: ...264 Clarus 590 GC User s Guide ...
Страница 279: ...Index ...
Страница 280: ...280 Clarus 590 GC User s Guide ...