102 . Clarus 590 GC User’s Guide
The Autosampler screen includes three tabs. The first two tabs display identical screens for setting
Program 1 and 2, allowing you to sequence two methods. The Tasks tab displays buttons to
perform tasks such as Park, Clean, and specify a priority injection. All of the screens provide status
information.
The method field allows the user to run the GC under the control of a method other than the
current active method.
Start Button
Touch the
Start
button to start a run. The Start button is active whether or not the system is ready.
Issuing the start command before the system is ready gives users the flexibility to set up and run a method
without having to wait for the system to attain ready status. As soon as you touch
OK
, the Start button
changes to
Stop
whether or not the system is ready. This enables you to cancel the run before it starts.
•
If the system is under external control (e.g. from TotalChrom) and a method has not yet been set up,
the Start Button is inactive.
•
In all other instances, even though the GC is not Ready, the Start Button will start a run.
•
When the GC oven starts to program, the Start Out signal will be asserted for 1 second.
•
The Ready Out signal will be asserted only when the GC is fully equilibrated and the GC has reached
full Ready status.
Stop Button
As soon as the
Start
button is touched, the
Stop
button appears on all top level screens.
Touching the Stop button before the system is running prompts the following dialog box:
Touch
OK
to cancel the run.
The status bar reflects that the system is running and the Start button changes to Stop.
Содержание CLARUS 590 GC
Страница 1: ...CLARUS 590 GC User s Guide GAS CHROMATOGRAPHY ...
Страница 9: ...About This Manual 9 Flame Ionization Detector FID 270 TCD 273 Autosampler 277 Index 279 ...
Страница 10: ...10 Clarus 590 GC User s Guide ...
Страница 11: ...1 Introduction ...
Страница 12: ...12 Clarus 590 GC User s Guide ...
Страница 21: ...2 Touch Screen Navigation ...
Страница 22: ...22 Clarus 590 GC User s Guide ...
Страница 48: ...48 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Страница 50: ...50 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Страница 59: ...3 Using the Active Method ...
Страница 60: ...60 Clarus 590 GC User s Guide ...
Страница 109: ...4 Setting Up the Detectors ...
Страница 110: ...110 Clarus 590 GC User s Guide ...
Страница 197: ...5 Using the Method Editor ...
Страница 198: ...198 Clarus 590 GC User s Guide ...
Страница 207: ...6 Using the Tools Menu ...
Страница 208: ...208 Clarus 590 GC User s Guide ...
Страница 224: ...224 Clarus 590 GC User s Guide ECD PID NPD FPD ...
Страница 225: ...Detectors 225 Output Configured ...
Страница 262: ...262 Clarus 590 GC User s Guide ...
Страница 263: ...7 Setting up a Typical Analysis ...
Страница 264: ...264 Clarus 590 GC User s Guide ...
Страница 279: ...Index ...
Страница 280: ...280 Clarus 590 GC User s Guide ...