90 . Clarus 590 GC User’s Guide
3.
Touch the
Relays
button.
The Relays screen appears:
4.
Open the split valve by touching the
Initial
check box for the
Split
field. This leaves the split vent
open.
NOTE: If you are using the Clarus GC for the first time, valves 1 and 2 will be OFF. Valves that are off are shown
as unchecked.
NOTE: Modifications to the valve settings become part of the Active Method and are automatically saved.
Measure and Adjust the Split Vent Flow
To measure and adjust the split vent flow:
1.
Open the detector cover (indicated below).
Содержание CLARUS 590 GC
Страница 1: ...CLARUS 590 GC User s Guide GAS CHROMATOGRAPHY ...
Страница 9: ...About This Manual 9 Flame Ionization Detector FID 270 TCD 273 Autosampler 277 Index 279 ...
Страница 10: ...10 Clarus 590 GC User s Guide ...
Страница 11: ...1 Introduction ...
Страница 12: ...12 Clarus 590 GC User s Guide ...
Страница 21: ...2 Touch Screen Navigation ...
Страница 22: ...22 Clarus 590 GC User s Guide ...
Страница 48: ...48 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Страница 50: ...50 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Страница 59: ...3 Using the Active Method ...
Страница 60: ...60 Clarus 590 GC User s Guide ...
Страница 109: ...4 Setting Up the Detectors ...
Страница 110: ...110 Clarus 590 GC User s Guide ...
Страница 197: ...5 Using the Method Editor ...
Страница 198: ...198 Clarus 590 GC User s Guide ...
Страница 207: ...6 Using the Tools Menu ...
Страница 208: ...208 Clarus 590 GC User s Guide ...
Страница 224: ...224 Clarus 590 GC User s Guide ECD PID NPD FPD ...
Страница 225: ...Detectors 225 Output Configured ...
Страница 262: ...262 Clarus 590 GC User s Guide ...
Страница 263: ...7 Setting up a Typical Analysis ...
Страница 264: ...264 Clarus 590 GC User s Guide ...
Страница 279: ...Index ...
Страница 280: ...280 Clarus 590 GC User s Guide ...