122 . Clarus 590 GC User’s Guide
Each pneumatic control consists of an outer knob (which is used to turn the hydrogen or air completely on
or off), a cap, and an inner gas flow adjustment screw.
Removing the cap exposes the gas flow adjustment screw. To set the air, flow use a small thin screwdriver
(P/N 09907273, supplied in the shipping kit) to turn the adjustment screw. Turn the screwdriver
counterclockwise to increase the flow or clockwise to decrease the flow. To set the hydrogen, use a hex
wrench (P/N 09907235, supplied in the shipping kit) to turn the adjustment screw. Turn the hex wrench
clockwise to increase the flow or counterclockwise to decrease the flow.
To adjust the hydrogen flow:
1.
Turn the air off by turning the outer ring of the air control valve completely clockwise.
2.
Turn on the hydrogen at the tank. Adjust the line pressure to 30 psig.
3.
Start the hydrogen flowing by turning the outer ring of the hydrogen pressure regulator completely
counterclockwise (see the previous figure).
NOTE: The outer knob is the on/off knob and it must be turned fully on for complete flow. In the case of the
pressure regulator, if this knob is only partially opened, the flow will be less than the setting made when
the knob was fully opened.
4.
To access the adjustment screw, remove the protective cap from the hydrogen pressure regulator.
5.
Insert the hex wrench (P/N 09907235) in the adjustment screw and turn it clockwise a few turns.
For safety reasons, the hydrogen pressure regulator is set to a very low flow before it is shipped
from the factory.
6.
From the System Status window touch
Tools
. From the Tools drop down menu touch
Utilities
.
Содержание CLARUS 590 GC
Страница 1: ...CLARUS 590 GC User s Guide GAS CHROMATOGRAPHY ...
Страница 9: ...About This Manual 9 Flame Ionization Detector FID 270 TCD 273 Autosampler 277 Index 279 ...
Страница 10: ...10 Clarus 590 GC User s Guide ...
Страница 11: ...1 Introduction ...
Страница 12: ...12 Clarus 590 GC User s Guide ...
Страница 21: ...2 Touch Screen Navigation ...
Страница 22: ...22 Clarus 590 GC User s Guide ...
Страница 48: ...48 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Страница 50: ...50 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Страница 59: ...3 Using the Active Method ...
Страница 60: ...60 Clarus 590 GC User s Guide ...
Страница 109: ...4 Setting Up the Detectors ...
Страница 110: ...110 Clarus 590 GC User s Guide ...
Страница 197: ...5 Using the Method Editor ...
Страница 198: ...198 Clarus 590 GC User s Guide ...
Страница 207: ...6 Using the Tools Menu ...
Страница 208: ...208 Clarus 590 GC User s Guide ...
Страница 224: ...224 Clarus 590 GC User s Guide ECD PID NPD FPD ...
Страница 225: ...Detectors 225 Output Configured ...
Страница 262: ...262 Clarus 590 GC User s Guide ...
Страница 263: ...7 Setting up a Typical Analysis ...
Страница 264: ...264 Clarus 590 GC User s Guide ...
Страница 279: ...Index ...
Страница 280: ...280 Clarus 590 GC User s Guide ...