Detectors . 47
Configuration Settings
The following graphic shows the
Configuration Screen
with the settings that appear on the
Active
Method Screen
.
Active Method
Configuration Screen
Manual Pneumatics
The manual pneumatics version of the FID shares several common elements with the PPC version. The
only difference is that the gas flow settings do not appear on the screen. The Auto-Ignite feature is not
available with manual pneumatics.
TCD
The Thermal Conductivity Detector (TCD) is a dual-channel detector which measures the difference in
thermal conductivity between carrier gas flowing through a reference channel and carrier gas plus sample
flowing through an analytical channel.
Four tungsten-rhenium filaments are connected in a bridge circuit, two in each channel. The TCD is set
up with the same amount of carrier gas flowing through both channels. The presence of a sample
component in the carrier gas flowing through the analytical channel changes the resistance of the
filaments in that channel. This unbalances the bridge and produces a signal proportional to the amount of
sample.
The following screens show the
Active Method
screen and the corresponding settings in the
Configuration
screen.
Содержание CLARUS 590 GC
Страница 1: ...CLARUS 590 GC User s Guide GAS CHROMATOGRAPHY ...
Страница 9: ...About This Manual 9 Flame Ionization Detector FID 270 TCD 273 Autosampler 277 Index 279 ...
Страница 10: ...10 Clarus 590 GC User s Guide ...
Страница 11: ...1 Introduction ...
Страница 12: ...12 Clarus 590 GC User s Guide ...
Страница 21: ...2 Touch Screen Navigation ...
Страница 22: ...22 Clarus 590 GC User s Guide ...
Страница 48: ...48 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Страница 50: ...50 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Страница 59: ...3 Using the Active Method ...
Страница 60: ...60 Clarus 590 GC User s Guide ...
Страница 109: ...4 Setting Up the Detectors ...
Страница 110: ...110 Clarus 590 GC User s Guide ...
Страница 197: ...5 Using the Method Editor ...
Страница 198: ...198 Clarus 590 GC User s Guide ...
Страница 207: ...6 Using the Tools Menu ...
Страница 208: ...208 Clarus 590 GC User s Guide ...
Страница 224: ...224 Clarus 590 GC User s Guide ECD PID NPD FPD ...
Страница 225: ...Detectors 225 Output Configured ...
Страница 262: ...262 Clarus 590 GC User s Guide ...
Страница 263: ...7 Setting up a Typical Analysis ...
Страница 264: ...264 Clarus 590 GC User s Guide ...
Страница 279: ...Index ...
Страница 280: ...280 Clarus 590 GC User s Guide ...