NPD Overview . 155
Adjust the Hydrogen Flow
NOTE: For safety reasons, the hydrogen pressure regulator is set in the off position (turned completely clockwise)
before it is shipped from the factory.
1.
Turn on the hydrogen gas at the tank and adjust the line pressure to 30 psig.
2.
Connect the tubing from a flow meter to the end of the hydrogen line at the NPD. It is marked
with a red sticker.
3.
Turn the hydrogen pressure regulator outer knob fully counterclockwise to open it.
NOTE: The outer knob is the on/off knob and it must be turned fully on for complete flow. In the case of the
pressure regulator, if this knob is only partially opened, the flow will be less than the setting made when
the knob was fully opened.
4.
To expose the adjustment screw, remove the protective cap from the hydrogen pressure regulator.
5.
Insert the hex wrench in the adjustment screw and turn it clockwise to increase the flow.
6.
From the System Status window touch
Tools
. From the Tools drop down menu touch
Utilities
.
7.
In the Utilities screen touch
Stopwatch
icon.
8.
Measure the flow rate. For best accuracy, use a soap bubble flowmeter volume that gives a reading
of at least 30 seconds.
Содержание CLARUS 590 GC
Страница 1: ...CLARUS 590 GC User s Guide GAS CHROMATOGRAPHY ...
Страница 9: ...About This Manual 9 Flame Ionization Detector FID 270 TCD 273 Autosampler 277 Index 279 ...
Страница 10: ...10 Clarus 590 GC User s Guide ...
Страница 11: ...1 Introduction ...
Страница 12: ...12 Clarus 590 GC User s Guide ...
Страница 21: ...2 Touch Screen Navigation ...
Страница 22: ...22 Clarus 590 GC User s Guide ...
Страница 48: ...48 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Страница 50: ...50 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Страница 59: ...3 Using the Active Method ...
Страница 60: ...60 Clarus 590 GC User s Guide ...
Страница 109: ...4 Setting Up the Detectors ...
Страница 110: ...110 Clarus 590 GC User s Guide ...
Страница 197: ...5 Using the Method Editor ...
Страница 198: ...198 Clarus 590 GC User s Guide ...
Страница 207: ...6 Using the Tools Menu ...
Страница 208: ...208 Clarus 590 GC User s Guide ...
Страница 224: ...224 Clarus 590 GC User s Guide ECD PID NPD FPD ...
Страница 225: ...Detectors 225 Output Configured ...
Страница 262: ...262 Clarus 590 GC User s Guide ...
Страница 263: ...7 Setting up a Typical Analysis ...
Страница 264: ...264 Clarus 590 GC User s Guide ...
Страница 279: ...Index ...
Страница 280: ...280 Clarus 590 GC User s Guide ...