Controlling the Autosampler . 107
The plunger speed during sample acquisition and injection will vary with the viscosity of the sample. To find
the optimum setting for the sample being analyzed, you should experiment with various syringe settings.
Highlighting the
Viscosity Control
Field use up and down arrow or keypad
buttons to enter
a value between 1 and 15. This value is the number of seconds the plunger pauses at the top-of-stroke
during sample acquisition.
NOTE: You can obtain improved precision with viscous samples by using the 0.5-mL syringe and septa in the
waste/wash vials in place of the diffusion caps.
Using Solvent Pre-Washes
Pre-washes refers to the number of solvent washes you want the system to perform prior to the pre-
injection sample washes. Highlight the
Solvent Pre-Wash
field and use the up and down arrow or
keypad
buttons to type in a value. Acceptable values are 0, 1 or 2. Pre-washes occur only at
the beginning of an autosampler program.
Priority Button
Press the
Priority
button to inject a priority sample during a run.
1.
Touch the Priority Button in the
Tasks
tab.
2.
In the Number of priority injection field use up and down arrow or keypad
buttons to
input the number of priority injections. Touch
Start
.
This option instructs the autosampler to interrupt the vial injection sequence and inject a "Priority"
vial that has been placed in the "Priority" position of the turntable. The priority vial will be injected
after completion of the current run.
After the priority vial has been injected (in accordance with the current program parameters), and
the priority run completed, the program continues with the next injection in sequence.
NOTE: Priority can only be selected when with the above screen is visible.
About Waste and Wash Vials
The Autosampler software has been optimized to use waste vials and solvent wash vials to minimize
contamination of the syringe needle.
Septa are available for the waste vials and solvent wash vials by ordering P/N N9302780. We recommend
using these septa in place of the diffuser caps when you are using:
•
Very viscous samples (to wipe the sample off the outside of the syringe needle).
•
Toxic solvents and samples.
•
Very volatile solvents.
Multi-Program Mode
Waste Vials
When operating in the Multi-Program Mode, the sample pre-washes are emptied into waste vial #1 for
program 1 and waste vial #3 for program 2. The solvent post-washes are emptied into waste vial #2 for
program 1 and #4 for program 2.
Содержание CLARUS 590 GC
Страница 1: ...CLARUS 590 GC User s Guide GAS CHROMATOGRAPHY ...
Страница 9: ...About This Manual 9 Flame Ionization Detector FID 270 TCD 273 Autosampler 277 Index 279 ...
Страница 10: ...10 Clarus 590 GC User s Guide ...
Страница 11: ...1 Introduction ...
Страница 12: ...12 Clarus 590 GC User s Guide ...
Страница 21: ...2 Touch Screen Navigation ...
Страница 22: ...22 Clarus 590 GC User s Guide ...
Страница 48: ...48 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Страница 50: ...50 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Страница 59: ...3 Using the Active Method ...
Страница 60: ...60 Clarus 590 GC User s Guide ...
Страница 109: ...4 Setting Up the Detectors ...
Страница 110: ...110 Clarus 590 GC User s Guide ...
Страница 197: ...5 Using the Method Editor ...
Страница 198: ...198 Clarus 590 GC User s Guide ...
Страница 207: ...6 Using the Tools Menu ...
Страница 208: ...208 Clarus 590 GC User s Guide ...
Страница 224: ...224 Clarus 590 GC User s Guide ECD PID NPD FPD ...
Страница 225: ...Detectors 225 Output Configured ...
Страница 262: ...262 Clarus 590 GC User s Guide ...
Страница 263: ...7 Setting up a Typical Analysis ...
Страница 264: ...264 Clarus 590 GC User s Guide ...
Страница 279: ...Index ...
Страница 280: ...280 Clarus 590 GC User s Guide ...