TCD Overview . 141
NOTE: Operating at too high a temperature can have an adverse effect on filament life.
Table 4-1. Bridge Current Range vs. Max. Detector Operating Temperature
Bridge Current Range
Recommended Operating Temperatures
for the Following Detectors
He or H
2
N
2
Argon
±
4
Up to 100
°
C
Do not Use 4
Do not Use 4
±
3
Up to 300
°
C
Do not Use 3
Do not Use 3
±
2
Up to 350
°
C
Up to 110
°
C
Do not Use 2
±
1
Up to 350
°
C
Up to 350
°
C
Up to 350
°
C
The correlation between bridge current and range is in Table 4-2:
Table 4-2. Correlation Between Bridge Current and Range
Bridge Current (mA)
Range Entry
Off
±0
40
±1
80
±2
120
±3
160
±4
NOTE: Entering a negative number reverses the detector polarity. For examples of reversing the polarity, see the
Clarus GC Hardware Guide, Maintenance Chapter, section “Practical Hints.”
Balance the Bridge
1.
In the System Status Screen touch the
Signal
icon
to get to the Signal screen. This will allow
you to observe the signal as you adjust the potentiometer.
2.
Allow the system to stabilize for about 20 min, then open the detector cover.
Содержание CLARUS 590 GC
Страница 1: ...CLARUS 590 GC User s Guide GAS CHROMATOGRAPHY ...
Страница 9: ...About This Manual 9 Flame Ionization Detector FID 270 TCD 273 Autosampler 277 Index 279 ...
Страница 10: ...10 Clarus 590 GC User s Guide ...
Страница 11: ...1 Introduction ...
Страница 12: ...12 Clarus 590 GC User s Guide ...
Страница 21: ...2 Touch Screen Navigation ...
Страница 22: ...22 Clarus 590 GC User s Guide ...
Страница 48: ...48 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Страница 50: ...50 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Страница 59: ...3 Using the Active Method ...
Страница 60: ...60 Clarus 590 GC User s Guide ...
Страница 109: ...4 Setting Up the Detectors ...
Страница 110: ...110 Clarus 590 GC User s Guide ...
Страница 197: ...5 Using the Method Editor ...
Страница 198: ...198 Clarus 590 GC User s Guide ...
Страница 207: ...6 Using the Tools Menu ...
Страница 208: ...208 Clarus 590 GC User s Guide ...
Страница 224: ...224 Clarus 590 GC User s Guide ECD PID NPD FPD ...
Страница 225: ...Detectors 225 Output Configured ...
Страница 262: ...262 Clarus 590 GC User s Guide ...
Страница 263: ...7 Setting up a Typical Analysis ...
Страница 264: ...264 Clarus 590 GC User s Guide ...
Страница 279: ...Index ...
Страница 280: ...280 Clarus 590 GC User s Guide ...