PPC Connect . 243
Calibrating the Carrier Gas Modules
CAUTION
ATTENTION
If you are performing this calibration procedure and you do not
have a column connected to the detector, you must plug the
detector fitting in the oven with a Swagelok 1/8-inch cap before
proceeding. This will ensure that all the gas flows up through
the detector and not into the oven.
Si vous effectuez cette procédure d'étalonnage et que vous
n'avez pas de colonne connectée au détecteur, vous devez
brancher le capot du détecteur au four avec un capuchon
Swagelok 1/8-inch avant de continuer. Cela garantira que tout le
gaz s'écoule à travers le détecteur et non dans le four.
If you are using a packed injector or programmed on-column injector, you can disconnect the column
and measure the flow at the exit of the injector.
If you are using a capillary injector or programmed split/splitless injector, you must remove the column.
Cap the injector exit in the oven and measure flows at the split vent exit.
To cap the injector exit in the oven, locate the Capillary Column Shipping Kit (P/N N6100160) and insert
the solid ferrule (P/N 09920511) into the 1/16-inch nut (P/N 09903392) and connect the nut to the exit of
the injector.
CAUTION
ATTENTION
To obtain the correct results, before you begin to calibrate you
must first Zero the Gas Modules.
Pour obtenir les résultats corrects, avant de commencer à
étalonner, vous devez d'abord zéro les modules de gaz.
1.
From the PPC screen touch the
PPC Configure
button. Touch
OK
on the Caution message.
Содержание CLARUS 590 GC
Страница 1: ...CLARUS 590 GC User s Guide GAS CHROMATOGRAPHY ...
Страница 9: ...About This Manual 9 Flame Ionization Detector FID 270 TCD 273 Autosampler 277 Index 279 ...
Страница 10: ...10 Clarus 590 GC User s Guide ...
Страница 11: ...1 Introduction ...
Страница 12: ...12 Clarus 590 GC User s Guide ...
Страница 21: ...2 Touch Screen Navigation ...
Страница 22: ...22 Clarus 590 GC User s Guide ...
Страница 48: ...48 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Страница 50: ...50 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Страница 59: ...3 Using the Active Method ...
Страница 60: ...60 Clarus 590 GC User s Guide ...
Страница 109: ...4 Setting Up the Detectors ...
Страница 110: ...110 Clarus 590 GC User s Guide ...
Страница 197: ...5 Using the Method Editor ...
Страница 198: ...198 Clarus 590 GC User s Guide ...
Страница 207: ...6 Using the Tools Menu ...
Страница 208: ...208 Clarus 590 GC User s Guide ...
Страница 224: ...224 Clarus 590 GC User s Guide ECD PID NPD FPD ...
Страница 225: ...Detectors 225 Output Configured ...
Страница 262: ...262 Clarus 590 GC User s Guide ...
Страница 263: ...7 Setting up a Typical Analysis ...
Страница 264: ...264 Clarus 590 GC User s Guide ...
Страница 279: ...Index ...
Страница 280: ...280 Clarus 590 GC User s Guide ...