FID Overview . 123
7.
Touch
Stopwatch
icon from the Utilities screen.
8.
Measure the flow rate. For best accuracy, use a soap bubble flowmeter volume that gives a reading
of at least 30 seconds.
NOTE: All gases exit through the top of the FID. If you use a flowmeter to measure the air and hydrogen gas
flows with a column connected to the FID and the carrier gas flowing, remember to subtract the carrier gas
flow from the total measured flow.
9.
Repeatedly adjust the hydrogen flow and measure the flow until you obtain a flow rate of about 30
mL/min.
NOTE: Both the carrier gas and hydrogen gas exit through the FID. For satisfactory operation of the FID, a flow
rate of 30 mL/min of hydrogen is required. If a column is connected to the FID, and the carrier gas is set
and flowing, you should add the carrier gas flow rate to the required hydrogen flow. For example, if the
carrier gas flow is 30 mL/min, adjust the hydrogen flow until the measured total flow rate (combined carrier
gas and hydrogen) is about 60 mL/min (30 mL/min of hydrogen plus 30 mL/min of carrier gas).
Содержание CLARUS 590 GC
Страница 1: ...CLARUS 590 GC User s Guide GAS CHROMATOGRAPHY ...
Страница 9: ...About This Manual 9 Flame Ionization Detector FID 270 TCD 273 Autosampler 277 Index 279 ...
Страница 10: ...10 Clarus 590 GC User s Guide ...
Страница 11: ...1 Introduction ...
Страница 12: ...12 Clarus 590 GC User s Guide ...
Страница 21: ...2 Touch Screen Navigation ...
Страница 22: ...22 Clarus 590 GC User s Guide ...
Страница 48: ...48 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Страница 50: ...50 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Страница 59: ...3 Using the Active Method ...
Страница 60: ...60 Clarus 590 GC User s Guide ...
Страница 109: ...4 Setting Up the Detectors ...
Страница 110: ...110 Clarus 590 GC User s Guide ...
Страница 197: ...5 Using the Method Editor ...
Страница 198: ...198 Clarus 590 GC User s Guide ...
Страница 207: ...6 Using the Tools Menu ...
Страница 208: ...208 Clarus 590 GC User s Guide ...
Страница 224: ...224 Clarus 590 GC User s Guide ECD PID NPD FPD ...
Страница 225: ...Detectors 225 Output Configured ...
Страница 262: ...262 Clarus 590 GC User s Guide ...
Страница 263: ...7 Setting up a Typical Analysis ...
Страница 264: ...264 Clarus 590 GC User s Guide ...
Страница 279: ...Index ...
Страница 280: ...280 Clarus 590 GC User s Guide ...