98 . Clarus 590 GC User’s Guide
6.
Touch
OK
.
The split flow will now turn off before the Clarus GC becomes ready so that at the time of injection
the split flow is off.
7.
Touch
Add
.
The Add Timed Events screen appears.
8.
From the Event drop down menu select
Split A
(if you have the injector configured for position A)
or Split B (if you have the injector configured for position B).
9.
Touch the
Time
dialog box and use the up and down arrow or keypad
buttons to
enter a positive time for the event (for example, 2.00).
NOTE: The "Time" refers to the time (in minutes) into the GC run that you want to automatically turn the valve on.
The time depends on your sample, solvent, column, etc. Typical time values are between 0.25 and 2.0
minutes. An optimum "Time" is the time that provides maximum area counts (indicated by your data
handling device) with minimum solvent effect (tailing).
10.
Touch the
Value
dialog box and use the up and down arrow or keypad
buttons s to
enter a value of 50 mL/min.
Содержание CLARUS 590 GC
Страница 1: ...CLARUS 590 GC User s Guide GAS CHROMATOGRAPHY ...
Страница 9: ...About This Manual 9 Flame Ionization Detector FID 270 TCD 273 Autosampler 277 Index 279 ...
Страница 10: ...10 Clarus 590 GC User s Guide ...
Страница 11: ...1 Introduction ...
Страница 12: ...12 Clarus 590 GC User s Guide ...
Страница 21: ...2 Touch Screen Navigation ...
Страница 22: ...22 Clarus 590 GC User s Guide ...
Страница 48: ...48 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Страница 50: ...50 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Страница 59: ...3 Using the Active Method ...
Страница 60: ...60 Clarus 590 GC User s Guide ...
Страница 109: ...4 Setting Up the Detectors ...
Страница 110: ...110 Clarus 590 GC User s Guide ...
Страница 197: ...5 Using the Method Editor ...
Страница 198: ...198 Clarus 590 GC User s Guide ...
Страница 207: ...6 Using the Tools Menu ...
Страница 208: ...208 Clarus 590 GC User s Guide ...
Страница 224: ...224 Clarus 590 GC User s Guide ECD PID NPD FPD ...
Страница 225: ...Detectors 225 Output Configured ...
Страница 262: ...262 Clarus 590 GC User s Guide ...
Страница 263: ...7 Setting up a Typical Analysis ...
Страница 264: ...264 Clarus 590 GC User s Guide ...
Страница 279: ...Index ...
Страница 280: ...280 Clarus 590 GC User s Guide ...