56
OPERATING PROCEDURE
Standard Program Refraction
7
Measure the cylindrical power.
1) Press
.
The cross cylinder lens axis changes.
2) Measure the cylindrical power with the
cross cylinder lens.
Follow the same procedure as 2) of Step 6.
On the chart, go to the next red-green test.
8
Refine the spherical power with the red-green test.
1) Press
.
The SPH +0.50 D lens is automatically
added to fog the vision. The red-green chart
is presented.
2) Reduce fogging gradually until the
sharpness of the letters on the red and
green sides appears equal.
Letters on the red side are sharper.
Turn
the dial clockwise one increment.
Letters on the green side are sharper.
Turn the dial counterclockwise one increment.
• When the patient cannot see the red and green sides equally, make the red side sharper
slightly. It avoids overcorrection.
9
Refine the spherical power to achieve the best possible visual acuity with the most plus
power.
1) Press
.
The isolated 20/20 line is presented.
2) Check the best possible visual acuity
with
or
.
3) Refine the spherical power to achieve
the best possible visual acuity with the
most plus power.
161221687130 承認済
Uncontrolled copy when printed.
Содержание RT-5100
Страница 12: ...X 161221687130 承認済 Uncontrolled copy when printed ...
Страница 18: ... XVI 161221687130 承認済 Uncontrolled copy when printed ...
Страница 46: ...28 BEFORE USE Getting Started and Exiting 161221687130 承認済 Uncontrolled copy when printed ...
Страница 226: ...208 TEST METHOD Visual Acuity Test by ETDRS chart 161221687130 承認済 Uncontrolled copy when printed ...
Страница 274: ...256 MAINTENANCE List of Replacement Parts 161221687130 承認済 Uncontrolled copy when printed ...
Страница 302: ...284 EMC ELECTROMAGNETIC COMPATIBILITY 161221687130 承認済 Uncontrolled copy when printed ...