ADVANCED FEATURES
Range of Clear Vision
3
4
After adjusting the power, press
to
return to the measurement screen.
The value adjusted with the Range of Clear
Vision test screen is set to [Fin*].
5
When checking the near power, press
.
• When the refractor head is converging, have the patient off their face temporarily from the
refracter head.
• The WD that can converge with the refractor main body is within range of 35 to 70 cm in 5
cm increments.
Therefore, convergence of the refractor main body is performed in 5 cm increments of WD conver-
sion.
Ex) When the WD is changed to 38 cm, the convergence and near PD is the same as for when
WD=40.
When WD is 35cm or less or 70cm or more, the convergence and near PD when WD is 35cm and
75cm are set to the main body.
The refractor head converges to the changed working distance.
6
If necessary, change values. If there are no problems, touch [OK].
To adopt the final prescription data before change, touch [OK] after touching [Final].