53
OPERATING PROCEDURE
Standard Program Refraction
2
2.5.2
Program Day/Night
Program
is a program to determine the day full correction value based on the objective measure-
ment data and then determine the night full correction value based on the day full correction value.
Pressing
during the program shows sample instructions to the patient. (The instruction may
not appear depending on the test. At this time, the normal help screen appears.)
• To measure with the program
, chart presenting device with night measurement
function (SSC-350, SSC-370, SC-1600, and SC-1600Pola) are necessary.
1
Enter the AR data.
1) Measure the patient’s eye with an auto refractometer and print the result.
2) Press
.
Confirm that the data No. is the same as the
print data No. of the auto refractometer.
See "2.2.1 From an auto refractometer"
(page 30).
3) Press
.
The AR data appears in the subjective and
AR data fields.
2
Select Program
.
When the program is not
, the program can be switched with
+
.