164
TEST METHOD
Binocular Visual Function Test
[Procedure example]
1
Present the binocular balance chart.
Press
,
,
,
, or
.
The polarizing filters are placed in the measuring windows. (The prisms are entered for the
SC-2000, SC-1600.)
The RT-5100 goes into SPH mode.
2
Fog the binocular vision.
Press
and turn the dial counterclockwise so that the binocular VA becomes between
20/30 and 20/25.
The VA of each letter on the chart is 20/50, 20/40, 20/30, 20/25, and 20/20 moving from the
left to right.
3
Ask the patient, “Which is sharper, the top or bottom line?"
The top line is sharper.
Press
and add SPH +0.25 D. (Turn the dial counterclock-
wise.)
The bottom line is sharper.
Press
and add SPH +0.25 D. (Turn the dial counterclock-
wise.)
The top and bottom lines should appear equal. When the patient cannot see them equally,
determine with which eye they view the binocular balance chart more clearly while referring
to the LM data or dominant eye.
4
Clear fogging applied in Step 2.
Press
and turn the dial.
161221687130 承認済
Uncontrolled copy when printed.
Содержание RT-5100
Страница 12: ...X 161221687130 承認済 Uncontrolled copy when printed ...
Страница 18: ... XVI 161221687130 承認済 Uncontrolled copy when printed ...
Страница 46: ...28 BEFORE USE Getting Started and Exiting 161221687130 承認済 Uncontrolled copy when printed ...
Страница 226: ...208 TEST METHOD Visual Acuity Test by ETDRS chart 161221687130 承認済 Uncontrolled copy when printed ...
Страница 274: ...256 MAINTENANCE List of Replacement Parts 161221687130 承認済 Uncontrolled copy when printed ...
Страница 302: ...284 EMC ELECTROMAGNETIC COMPATIBILITY 161221687130 承認済 Uncontrolled copy when printed ...