39
OPERATING PROCEDURE
Entering Data
2
When calling up data from the ARK-10000 and OPD-Scan III via Eye Care card
The optional Eye Care card allows day and night data to be entered without connecting the ARK-
10000 and OPD-Scan III directly to the RT-5100.
When entering day data only, see "2.2.4 From an Eye Care card" (page 34).
* For detailed procedure for transferring data from the ARK-10000 or OPD-Scan III to the Eye Care card, refer to
the ARK-10000 or OPD-Scan III Operator’s Manual.
• Set the “AR port” parameter or the “LM port” parameter to “WF(OPD)”.
• On the ARK-10000 Settings screen, check “Send HD Exam data” and “Use Eye Care
Card”.
• On the Communication tab in the OPD-Scan III Settings screen, set “Connection Device” to
“EyeCareCard” then check “Send HD Exam data”. On the Settings screen in the
Measurement tab of OPD-Scan III, check “RS232C connection” in the Option field.
1
Measure the patient’s eye with the ARK-10000 in ARK/CT mode. Write the measured
data to the Eye Care card. Or measure the patient’s eye with the OPD-Scan III in OPD/
CT mode and write the measured data to the Eye Care card.
For details, refer to the ARK-10000 or OPD-Scan III Operator’s Manual.
2
Insert the Eye Care card into the Eye Care card reader of the control box.
The measured data is automatically read from the Eye Care card to the control box. The day
data is entered in the AR field.
The data in the Eye Care card is cleared.
The refractor head is automatically set depending on the day data. The subjective refine-
ment starts.
is displayed on the left of the measurement mode display as the indication of day data.
When the day data is the Wave Front data, the program becomes the program set in the
“Program for WF data imported from OPD” parameter automatically. When the day data is
the AR median value, the program becomes the program set in the “Program for AR data
imported from OPD” parameter automatically. If either program is running, the program can-
not be switched when the Wave Front data is entered.
*To enter and use the Wave Front data, rewrite the program set in the “Program for WF data
imported from OPD” parameter so that it is compatible with the Wave Front data.
• When the “Data clear confirmation by EyeCare card insertion” parameter is set to Yes, if
data is read from Eye Care card with the measurement data displayed, a window is
displayed asking whether or not to read the Eye care card data after clearing the displayed
data.
Clear + input
Clear all of the displayed data and read the data in the Eye Care card.
Overwrite
Overwrite with the data in the Eye Care card without clearing the data displayed. Any
items in the displayed data that are not included in the Eye Care card data remain as they are.
(However, for the Subjective field, follow the “Preset power of Subj 1” parameter setting.)
3
Perform the subjective refinement by the same procedure as Steps 7 and later in
“
When calling up data from the ARK-10000 connected with a cable”.
161221687130 承認済
Uncontrolled copy when printed.
Содержание RT-5100
Страница 12: ...X 161221687130 承認済 Uncontrolled copy when printed ...
Страница 18: ... XVI 161221687130 承認済 Uncontrolled copy when printed ...
Страница 46: ...28 BEFORE USE Getting Started and Exiting 161221687130 承認済 Uncontrolled copy when printed ...
Страница 226: ...208 TEST METHOD Visual Acuity Test by ETDRS chart 161221687130 承認済 Uncontrolled copy when printed ...
Страница 274: ...256 MAINTENANCE List of Replacement Parts 161221687130 承認済 Uncontrolled copy when printed ...
Страница 302: ...284 EMC ELECTROMAGNETIC COMPATIBILITY 161221687130 承認済 Uncontrolled copy when printed ...