62
OPERATING PROCEDURE
Standard Program Refraction
2) Measure the cylindrical axis with the cross cylinder lens.
Flipped cross cylinder:
Ask the patient which is sharper, chart 1
presented by pressing
or chart 2 pre-
sented by pressing
Types of cross cylinder lens
When chart 1 is sharper.
Turn the dial counterclockwise one incre-
ment.
(Also possible by pressing
)
When chart 2 is sharper.
Turn the dial clockwise one increment.
(Also possible by pressing
)
Repeat the above until both charts appear
equal.
Auto cross cylinder:
Charts 1 and 2 are presented simultane-
ously. Both chart positions appear in the
lower left corner of the screen.
When chart 1 is sharper.
Turn the dial counterclockwise one incre-
ment.
(Also possible by pressing
)
When chart 2 is sharper.
Turn the dial clockwise one increment.
(Also possible by pressing
.)
Repeat the above until both charts appear
equal.
24
Measure the cylindrical power.
1) Press
.
The cross cylinder lens axis changes.
2) Measure the cylindrical power with the
cross cylinder lens.
Follow the same procedure as 2) of Step 23.
161221687130 承認済
Uncontrolled copy when printed.
Содержание RT-5100
Страница 12: ...X 161221687130 承認済 Uncontrolled copy when printed ...
Страница 18: ... XVI 161221687130 承認済 Uncontrolled copy when printed ...
Страница 46: ...28 BEFORE USE Getting Started and Exiting 161221687130 承認済 Uncontrolled copy when printed ...
Страница 226: ...208 TEST METHOD Visual Acuity Test by ETDRS chart 161221687130 承認済 Uncontrolled copy when printed ...
Страница 274: ...256 MAINTENANCE List of Replacement Parts 161221687130 承認済 Uncontrolled copy when printed ...
Страница 302: ...284 EMC ELECTROMAGNETIC COMPATIBILITY 161221687130 承認済 Uncontrolled copy when printed ...