6
BEFORE USE
Configuration
Display
(A) Measurement screen
Measurement mode
1. PD display and change key 2. Auxiliary lenses key
3. Far or near mode key
Time
Operation message
Program message
Function buttons
Present chart
4. Chart buttons
5. Current data
Next chart
6. Chart / 21 Point
Exams switching button
(only for Plus Package)
Date
7. Display data setting button 1
9. Display data setting button 2
8. Sub window 1
10. Sub window 2
1. PD display and change key
Enters the mode changing the pupillary distance.
Monocular PD (1/2 PD)
Enter by pressing
or
in PD mode.
Leave the mode
Press the PD key again.
2. Auxiliary lenses key
Displays the current inserted auxiliary lens.
Pressing this key displays the Auxiliary selection screen.
3. Far or near mode key
Displays the current measurement mode (far or near mode).
When in far mode, the key indication is “FAR” and when in near mode, the key indication is “NEAR”.
The background of the data turns white when in far vision measurement mode and yellow when in near vision
measurement mode. Far mode is automatically selected when the power is turned on or after data is cleared.
Near mode is used to test the near visual function after far vision testing and addition power measurement. The
refractor head converges to the working distance of 40 cm (variable) automatically in near mode.
The ADD key is not available in near mode. The near SPH values are calculated by adding the ADD values to the
far SPH values. The appropriate lenses are placed. Only the far SPH values in near mode can also be used when
the “SPH Far
Near” parameter is set to “SPH”.
4. Chart buttons
Switches the chart.
5. Current data
Displays the data of the lens that is currently inserted into the refractor. Press other data to be changed as
desired.
The numeric field that can be changed is highlighted.
The value can be changed with the dial,
or
.
The currently set power is displayed in black.
ADD and Prism value become black when the value is set in ADD measurement mode or Prism measurement
mode.
161221687130 承認済
Uncontrolled copy when printed.
Содержание RT-5100
Страница 12: ...X 161221687130 承認済 Uncontrolled copy when printed ...
Страница 18: ... XVI 161221687130 承認済 Uncontrolled copy when printed ...
Страница 46: ...28 BEFORE USE Getting Started and Exiting 161221687130 承認済 Uncontrolled copy when printed ...
Страница 226: ...208 TEST METHOD Visual Acuity Test by ETDRS chart 161221687130 承認済 Uncontrolled copy when printed ...
Страница 274: ...256 MAINTENANCE List of Replacement Parts 161221687130 承認済 Uncontrolled copy when printed ...
Страница 302: ...284 EMC ELECTROMAGNETIC COMPATIBILITY 161221687130 承認済 Uncontrolled copy when printed ...