![background image](http://html1.mh-extra.com/html/nidek-medical/rt-5100/rt-5100_operators-manual_1688048161.webp)
143
ADVANCED FEATURES
Parameter Settings
3
Display the Day/Night switching button: Yes, No
Factory setting: No
Selection of whether or not to input the day/night data. This parameter can be selected only when the
RT is connected with a chart presenting device equipped with the night vision test function.
Program for WF data imported from OPD: A, B, C, D, E,
Factory setting: B
Setting for the program to automatically be set when connected to the ARK-10000 (V2.11 or later) or
OPD-Scan III and the Wave Front value is entered, with the “AR port” or “LM port” parameters set to
“WF(OPD)”.
The program
appears only when the RT is connected with a chart presenting device equipped
with the night vision test function.
Program for AR data imported from OPD: A, B, C, D, E,
Factory setting: A
Setting for the program to automatically be set when connected to the ARK-10000 (V2.11 or later) or
OPD-Scan III and the AR median value is entered, with the “AR port” or “LM port” parameters set to
“WF(OPD)”.
The program
appears only when the RT is connected with a chart presenting device equipped
with the night vision test function.
Program for Night data imported from OPD: A, B, C, D, E,
Factory setting: A
Setting for the program to automatically be set when night data is entered from the connected ARK-
10000 (V2.13 or later) or OPD-Scan III and the day and night data are entered at the same time, with
the “AR port” or “LM port” parameters set to “WF(OPD)”.
The program
appears only when the RT is connected with a chart presenting device equipped
with the night vision test function.
Data clear confirmation by EyeCare card insertion: Yes, No
Factory setting: No
Selection of whether to ask if data is entered after the data that is being measured is cleared when
data is entered from the Eye Care card to the RG-5100.
Write to Eye Care card: Yes, No
Factory setting: No
Yes
When printing or when the Eye Care card is inserted, the examination data for one person is
written to the Eye Care card.
No
Does not write to Eye Care card.
Input sequence from PC: Hold, Pass
Factory setting: Hold
Selection of whether to hold or overwrite the data imported from a computer (PC) when the RT-5100
already has data.
Hold
Holds the data imported from the PC temporarily.
Press
and
or
to enter and overwrite the data.
Pass
Overwrites and clears the data already displayed in the RT-5100 when another data is
imported from the PC.
* When the RT-5100 has no data, it functions as the same as Pass regardless of this parameter
setting.
161221687130 承認済
Uncontrolled copy when printed.
Содержание RT-5100
Страница 12: ...X 161221687130 承認済 Uncontrolled copy when printed ...
Страница 18: ... XVI 161221687130 承認済 Uncontrolled copy when printed ...
Страница 46: ...28 BEFORE USE Getting Started and Exiting 161221687130 承認済 Uncontrolled copy when printed ...
Страница 226: ...208 TEST METHOD Visual Acuity Test by ETDRS chart 161221687130 承認済 Uncontrolled copy when printed ...
Страница 274: ...256 MAINTENANCE List of Replacement Parts 161221687130 承認済 Uncontrolled copy when printed ...
Страница 302: ...284 EMC ELECTROMAGNETIC COMPATIBILITY 161221687130 承認済 Uncontrolled copy when printed ...