V
During Use
WARNING
•
When moving the refractor head, make sure that there are no obstacles in
its path.
If the arm is swung carelessly, it may bump against someone or something and injury
or malfunction may occur. In addition, the near point rod may topple over and injury
may occur.
CAUTION
•
Unplug the power cord and contact NIDEK or your authorized distributor if
the internal wires are exposed, the table turns on or off when the power
cord is moved, or the cord and/or plug are not to the touch.
This may result in electric shock or fire.
•
Do not look directly into the illumination for the near vision test or VD
check.
The retina of the operator or patient may be damaged.
•
Keep the measuring windows free of fingerprints and smudges.
Failure to do so could affect the data accuracy.
•
In the event of smoke or strange odors, immediately turn off the device and
disconnect the power plug from the outlet. Once it is determined that the
smoke has stopped, contact NIDEK or your authorized distributor.
Continued use of the device under such abnormal conditions may cause fire or elec-
tric shock.
•
When the device is not in use, turn it off and cover it with its dust cover.
If the device is not covered for an extended period, the accumulation of dust may
affect the data accuracy.
•
Do not touch the touch-screen panel with anything other than the tip of a
touch-screen pen.
Contact with a hard or sharp object such as a ball point pen may scratch the panel.
In addition, although the tip of a touch-screen pen is covered with resin which is
unlikely to scratch the panel, the panel may be scratched by strongly pressing the pen
against the panel.
Do not touch the panel with fingers. Doing so could decrease the panel sensitivity or
make the panel dirty. It may affect the test accuracy.
•
Perform the visual and operation checks before using the device. If any
abnormality is found, do not use the device.
Continued use of the device under such abnormal conditions may affect the data
accuracy. Unexpected malfunction or faulty diagnosis may induce unexpected health
hazards.
•
Before each patient, always clean the patient’s contact area (forehead rest
and face shields) using disinfectant alcohol.
161221687130 承認済
Uncontrolled copy when printed.
Содержание RT-5100
Страница 12: ...X 161221687130 承認済 Uncontrolled copy when printed ...
Страница 18: ... XVI 161221687130 承認済 Uncontrolled copy when printed ...
Страница 46: ...28 BEFORE USE Getting Started and Exiting 161221687130 承認済 Uncontrolled copy when printed ...
Страница 226: ...208 TEST METHOD Visual Acuity Test by ETDRS chart 161221687130 承認済 Uncontrolled copy when printed ...
Страница 274: ...256 MAINTENANCE List of Replacement Parts 161221687130 承認済 Uncontrolled copy when printed ...
Страница 302: ...284 EMC ELECTROMAGNETIC COMPATIBILITY 161221687130 承認済 Uncontrolled copy when printed ...