155
TEST METHOD
Visual Acuity Test
4
4.1.6
VA measurement of day data in brightness for Night mode
Measure VA in night mode with the day corrective lens inserted. This data can be used as reference
when the day glasses are used in night.
• To measure the day data in night brightness, a chart presenting device with Night
measurement function (SSC-350/SSC-370/SC-1600/SC-1600Pola) is necessary.
1
Press
.
The chart becomes night mode.
The measurement screen switches from
VA mode to “L-illu.VA" and becomes VA
measurement mode in night brightness.
2
Measure VA for right eye, left eye, and both
eyes.
Press
/
/
and turn the dial to
input the VA value.
The VA value can be input either with
/
.
3
VA measurement mode in night brightness can be left by switching to the other modes
(S/C/A etc.)
161221687130 承認済
Uncontrolled copy when printed.