37
OPERATING PROCEDURE
Entering Data
2
6
Enter the data.
Press
.
The refractor head is automatically set depend-
ing on the day data. The subjective refinement
starts.
is displayed next to the Measurement mode
display as the indication of day data.
When the day data is the Wave Front data, the
program becomes the program set in the “Pro-
gram for WF data imported from OPD” parame-
ter automatically. When the day data is the AR median value, the program becomes the
program set in the “Program for AR data imported from OPD” parameter automatically. If
either program is running, the program cannot be switched when the Wave Front data is
entered.
*To enter and use the Wave Front data, rewrite the program set in the “Program for WF data
imported from OPD” so that it is compatible with the Wave Front data.
7
Perform the subjective refinement based on the day data.
8
Press the
function button.
The refractor head is automatically set depending on the night data. The subjective refine-
ment starts.
is displayed next to the Measurement mode display as the indication of night data.
The program automatically switches to that set by the “Program for Night data imported from
OPD” parameter.
If the program is running, the program cannot be switched when the night data is entered.
9
Prepare an environment suitable for night measurement. Perform the subjective refine-
ment based on the night data and print the measured result.
For day data, “For day” is indicated on the printout, showing that the data is for glasses used
in bright environments. See “
When printing day/night data (page 89)”.
For night data, “For night” is indicated on the printout, showing that the data is for glasses
used in dark environments.
The data is cleared.
161221687130 承認済
Uncontrolled copy when printed.
Содержание RT-5100
Страница 12: ...X 161221687130 承認済 Uncontrolled copy when printed ...
Страница 18: ... XVI 161221687130 承認済 Uncontrolled copy when printed ...
Страница 46: ...28 BEFORE USE Getting Started and Exiting 161221687130 承認済 Uncontrolled copy when printed ...
Страница 226: ...208 TEST METHOD Visual Acuity Test by ETDRS chart 161221687130 承認済 Uncontrolled copy when printed ...
Страница 274: ...256 MAINTENANCE List of Replacement Parts 161221687130 承認済 Uncontrolled copy when printed ...
Страница 302: ...284 EMC ELECTROMAGNETIC COMPATIBILITY 161221687130 承認済 Uncontrolled copy when printed ...