35
OPERATING PROCEDURE
Entering Data
2
2.2.5
Entering day and night data
When the “Display the Day/Night switching button” parameter is set to “Yes”,
is displayed
and night data can be input separately from day data.
* Day data is measurement data for bright environment and night data is measurement data for dark
envionment.
• To measure in Night mode, a chart presenting device with Night measurement function
(SSC-350/SSC-370/SC-1600/SC-1600Pola) is necessary.
By pressing
, Day/Night mode can be alternated for data input.
Day mode
Night mode
Day mode and Night mode are indicated by (
/
) in the current mode field.
When no data is input for the specified mode, the S/C/A values displayed before pressing
are copied. However, when the subjective mode in which no data input has been specified, the “Preset
power of Subj 1” parameter setting is prioritized. (For the Day mode, the day AR value or LM value is
copied. For the Night mode, the night AR value or LM value is copied.)
In addition, when starting or the displayed data is cleared, the mode is Day mode.
Switching among Unaided and Final in Day/Night mode
Pressing
,
,
,
, or
in Day mode or Night mode switches the data selec-
tion mode from among those Unaided to Final modes.
When nothing is input for the specified mode, the S/C/A values displayed before pressing are copied.
However, when the subjective mode whose data is not input is specified, the “Subjective start value 1”
parameter setting is prioritized.
Input data to Night mode from AutoRef/Lensmeter
In Night mode, input data from an auto refractometer or lensmeter according to the procedures in
"2.2.1 From an auto refractometer" (page 30) and "2.2.2 From a lensmeter" (page 32).
161221687130 承認済
Uncontrolled copy when printed.
Содержание RT-5100
Страница 12: ...X 161221687130 承認済 Uncontrolled copy when printed ...
Страница 18: ... XVI 161221687130 承認済 Uncontrolled copy when printed ...
Страница 46: ...28 BEFORE USE Getting Started and Exiting 161221687130 承認済 Uncontrolled copy when printed ...
Страница 226: ...208 TEST METHOD Visual Acuity Test by ETDRS chart 161221687130 承認済 Uncontrolled copy when printed ...
Страница 274: ...256 MAINTENANCE List of Replacement Parts 161221687130 承認済 Uncontrolled copy when printed ...
Страница 302: ...284 EMC ELECTROMAGNETIC COMPATIBILITY 161221687130 承認済 Uncontrolled copy when printed ...