230
21 Point Exams (only for Plus Package)
21 Point Exams (only for Plus Package)
5.2.16 #12B Supravergence (Far), #18B Supravergence (Near)
Setting contents:
#12B Supra(F)
#18B Supra(N)
Use chart
Far point horizontal line chart
For the model other than CP series, SSC
series, SC series U and M, the chart is
horizontal line masked VA chart.
Near point horizontal line chart
Set the near point chart II C (VA0.63) or D
(VA1.0) manually.
Auxiliary lens
(R/L)
Binocular open (rotary prism)
Binocular open (rotary prism)
Set power
Subj (F) + #8 H.Phoria (F)
Subj (N) + #13 H.Phoria (N)
* The RT-5100 is set to add prism to the right eye as default. Therefore, right supravergence
test is set as a default. Setting the “Prism for Supra. and Infra.” parameter to “L” adds prism
to the left eye and sets to the left supravergence test. However, only one test can be stored.
[Procedure example]
The following describes the right superavergence test. When setting the Prism for Supra.
and Infra. to “L” and performing the left supravergence, the dial rotation is reversed.
1
With #7 Subjective value entered, press [#12B Supra(F)] or [#18B Supra(N)].
For the initial state, horizontal prism corresponding to the horizontal phoria value is set. If the
horizontal prism is not necessary, press [H.P] to clear the prism.
For [#18B Supra(N)], set the near point chart II C (VA0.63) or D (VA1.0) manually at the 40
cm position.
[#12B Supra(F)]
[#18B Supra(N)]
Operation message
2
According to the operation message on the left bottom of the screen, turn the dial coun-
terclockwise to add the BD prism powers to the single eye until the chart appears disso-
ciated.
*35
Also possible by pressing
.
*35.
Turning the dial changes the value in 0.2
increments (0.1
for single eye). Turning the dial with pressing [Shift] changes
the value in 1.0
increments (0.5
for single eye).
161221687130 承認済
Uncontrolled copy when printed.
Содержание RT-5100
Страница 12: ...X 161221687130 承認済 Uncontrolled copy when printed ...
Страница 18: ... XVI 161221687130 承認済 Uncontrolled copy when printed ...
Страница 46: ...28 BEFORE USE Getting Started and Exiting 161221687130 承認済 Uncontrolled copy when printed ...
Страница 226: ...208 TEST METHOD Visual Acuity Test by ETDRS chart 161221687130 承認済 Uncontrolled copy when printed ...
Страница 274: ...256 MAINTENANCE List of Replacement Parts 161221687130 承認済 Uncontrolled copy when printed ...
Страница 302: ...284 EMC ELECTROMAGNETIC COMPATIBILITY 161221687130 承認済 Uncontrolled copy when printed ...