136
ADVANCED FEATURES
Parameter Settings
Fog for binocular balance: Yes, No
Factory setting: Yes
Selection of whether or not to fog the vision automatically so that the patient’s visual acuity become 20/
30 to 20/20.
VA format: Normal
Factory setting: Normal
Selection of the VA value display and print format.
Now, the option is Normal only.
Preset VA: 1 letter, H.line, No
Factory setting: H.line
Selection of whether to isolate a single letter or horizontal line on the letter chart whose visual acuity is
estimated based on the AR or LM data. However, the estimated visual acuity chart is not presented
when the data is entered in the final field.
1 letter
Displays the estimated visual acuity chart with a single letter isolation.
H. line
Displays the estimated visual acuity chart with a horizontal line isolation.
No
The estimated visual acuity chart is not presented.
Preset ADD: Setting 1 (Low), Setting 2 (High), None
Factory setting: Setting 1 (Low)
Selection of whether or not to enter the patient’s age and preset the estimated additional power when
,
, or
is pressed.
For the additional power to be preset, see "8.3 VA Conversion Table" (page 267).
SPH Far
Near: SPH, SPH + ADD
Factory setting: SPH + ADD
Setting for the sphere values when far mode is switched to near one.
SPH
The sphere values in far mode are also used in near mode.
SPH + ADD
The additional powers are added to the sphere values in far mode.
(–) ADD: Yes, No
Factory setting: No
Selection of whether or not to enter the negative value as additional power.
Working distance (WD): 35 cm to 70 cm
Factory setting: 40 cm
Setting for the working distance in 5 cm increments.
NPC calculation: Detail, Simple
Factory setting: Detail
For the near point of convergence, when calculating the meter angle (MA), setting for whether to
include the distance between the base of patient’s nose (the inner side of the lens) and the cycloduc-
tion (0.012 + 0.013 =0.025).
Detail
calculate including 0.025
Simple
calculate excluding 0.025
161221687130 承認済
Uncontrolled copy when printed.
Содержание RT-5100
Страница 12: ...X 161221687130 承認済 Uncontrolled copy when printed ...
Страница 18: ... XVI 161221687130 承認済 Uncontrolled copy when printed ...
Страница 46: ...28 BEFORE USE Getting Started and Exiting 161221687130 承認済 Uncontrolled copy when printed ...
Страница 226: ...208 TEST METHOD Visual Acuity Test by ETDRS chart 161221687130 承認済 Uncontrolled copy when printed ...
Страница 274: ...256 MAINTENANCE List of Replacement Parts 161221687130 承認済 Uncontrolled copy when printed ...
Страница 302: ...284 EMC ELECTROMAGNETIC COMPATIBILITY 161221687130 承認済 Uncontrolled copy when printed ...