OPERATING PROCEDURE
Standard Program Refraction
2
2.5
Standard Program Refraction
The RT-5100 includes two standard programs (Programs A and B) for full-corrected far value and pre-
scription. See "8.8 Standard Programs" (page 272) for all test items in the standard programs.
For Plus Package, standard program #7 obtaining full-corrected far value is equipped other than pro-
grams A and B.
In this manual, standard program A and standard program #7 are presented as an example.
2.5.1
Program A
Containing unaided and aided visual acuity tests, this program uses the Final Fit (Auto adjustment
function) to automatically adjust a prescription to suit the patient.
Pressing
during the program shows sample instructions to the patient. (The instruction may
not appear depending on the test. At this time, the normal help screen appears.)
1
Select Program A.
The program is switched with
+
.
2
Press
.
Program A starts.
3
Enter the AR data.
*9
1) Measure the patient’s eye with an auto
refractometer and print the result.
2) Press
.
Confirm that the data No. is the same as the
print data No. of the auto refractometer.
See "2.2.1 From an auto refractometer"
(page 30).
*9.
When AR data and LM data have been entered before standard program A refraction is started, Steps 3 and Steps 7 to 8
are automatically skipped.
161221687130 承認済
Uncontrolled copy when printed.
Содержание RT-5100
Страница 12: ...X 161221687130 承認済 Uncontrolled copy when printed ...
Страница 18: ... XVI 161221687130 承認済 Uncontrolled copy when printed ...
Страница 46: ...28 BEFORE USE Getting Started and Exiting 161221687130 承認済 Uncontrolled copy when printed ...
Страница 226: ...208 TEST METHOD Visual Acuity Test by ETDRS chart 161221687130 承認済 Uncontrolled copy when printed ...
Страница 274: ...256 MAINTENANCE List of Replacement Parts 161221687130 承認済 Uncontrolled copy when printed ...
Страница 302: ...284 EMC ELECTROMAGNETIC COMPATIBILITY 161221687130 承認済 Uncontrolled copy when printed ...