114
ADVANCED FEATURES
Setting Auxiliary Lenses
[6
U]: Base up prism of 6
[10
I]: Base in prism of 10
[
]: Right eye: Horizontal maddox rod, Left eye: Open aperture
[
]: Right eye: Open aperture, Left eye: Vertical maddox rod
[
]: Pinhole plate (2 mm in diameter)
=
?
: Break prism of binocular balance
The prism value can be changed with the following [+] and [-].
The value can be changed for both eyes at the same time.
Changeable range: R 3.0
BD/L 3.0
BU to R 10.0
BD/L 10.0
BU
=
?㧦
Break prism of horizontal phoria
The prism value can be changed with the following [+] and [-].
The value can be changed for both eyes at the same time.
Changeable range: R 3.0
BU/L 3.0
BD to R 10.0
BU/L 1.0
BD
=
?㧦
Break prism of vertical phoria
The prism value can be changed with the following [+] and [-].
The value can be changed for both eyes at the same time.
Changeable range: R 5.0
BI/L 5.0
BI to R 15.0
BI/L 15.0
BI
=
?㧦
Break prism of #15A horizontal phoria (according to #14A value) (only for Plus
Package)
The prism value can be changed with the following [+] and [-].
The value can be changed for both eyes at the same time.
BU
BD Changeable range: R 3.0
BU/L 3.0
BD to R 10.0
BU/L 10.0
BD
161221687130 承認済
Uncontrolled copy when printed.
Содержание RT-5100
Страница 12: ...X 161221687130 承認済 Uncontrolled copy when printed ...
Страница 18: ... XVI 161221687130 承認済 Uncontrolled copy when printed ...
Страница 46: ...28 BEFORE USE Getting Started and Exiting 161221687130 承認済 Uncontrolled copy when printed ...
Страница 226: ...208 TEST METHOD Visual Acuity Test by ETDRS chart 161221687130 承認済 Uncontrolled copy when printed ...
Страница 274: ...256 MAINTENANCE List of Replacement Parts 161221687130 承認済 Uncontrolled copy when printed ...
Страница 302: ...284 EMC ELECTROMAGNETIC COMPATIBILITY 161221687130 承認済 Uncontrolled copy when printed ...