• The rotary prism lenses are removed from the measuring windows when the already
selected button is pressed again in prism mode.
ex.When
is pressed in the base in/out prism, the rotary prism lenses are removed.
Appearance of
chart
Phoria
Correction
<1> The upper col-
umn is on the left.
Esophoria
Turn the dial clockwise to add the BO prism powers until
the two columns line up.
<2> The upper col-
umn is on the right
Exophoria
Turn the dial counterclockwise to add the BI prism
powers until the two columns line up.
212
21 Point Exams (only for Plus Package)
21 Point Exams (only for Plus Package)
*Turning the dial changes the value in 0.5
increments for single eye.
*Use
or
instead of the dial for changing the prism value in 0.1
increments for sin-
gle eye.