134
ADVANCED FEATURES
Parameter Settings
Detailed parameter options
SPH step: 0.12 D, 0.25 D
Factory setting: 0.25 D
Setting for the increment to adjust sphere values. Selectable between 0.12 D and 0.25 D.
CYL mode: –, +
Factory setting:
–
Setting for the cylinder reading. Selectable b and –.
AXIS step: 1°, 5°
Factory setting: 5
°
Setting for the increment to adjust axis values. Selectable between 1
°
and 5
°
.
SPH step (Shift): 0.50 D to 3.00 D
Factory setting: 1.00 D
Setting for the increment to adjust sphere values with
.
CYL step (Shift): 1.00 D, 2.00 D, 3.00 D
Factory setting: 1.00 D
Setting for the increment to adjust cylinder values with
. Selectable among 1.00 D, 2.00 D, and
3.00 D.
AXIS step (Shift): 1°
5°/5°
1°, 1°
5°/5°
15°
Factory setting: 1°
5°/5°
1°
Setting for the increment to adjust axis values with
when the AXIS step parameter is set to 5.
1°
5°/5°
1°
In increments of 1° while pressing
1°
5°/5°
15°
In increments of 15° while pressing
When the AXIS step parameter is set to 1°, the increment to adjust axis values with
is 5°.
AXIS auto step: 0.00 D to 4.00 D
Factory setting: 0.00 D
In cylindrical axis measurement with a cross cylinder lens, the AXIS step becomes 5° when the
obtained cylinder value is less than the setting. The AXIS step becomes 1° when the obtained cylinder
value is the setting one or more. The setting value changes in 0.25 D increments.
When the setting is 0.00 D, the AXIS step does not change regardless of the cylinder value.
Indication of Axis 0°: 0°, 180°
Factory setting: 180
°
Selection of whether to display “0
°
” or “180
°
” when the cylinder axis is in the horizontal direction. How-
ever, when the cylindrical power is 0 D, the cylinder axis is displayed as 0
°
.
180
°
When cylindrical power and the cylinder axis are horizontal, the cylinder axis is displayed
as 180
°
and printed.
0
°
When the cylindrical power and when the cylinder axis is displayed as 0
°
and printed.
Cross Cylinder (XC) test: AUTO, ±0.25D, ±0.50D, ±0.25D/
○
Factory setting: AUTO
Setting for the cross cylinder lens placed in the cross cylinder test.
AUTO
±0.25 D Auto cross cylinder lens
161221687130 承認済
Uncontrolled copy when printed.
Содержание RT-5100
Страница 12: ...X 161221687130 承認済 Uncontrolled copy when printed ...
Страница 18: ... XVI 161221687130 承認済 Uncontrolled copy when printed ...
Страница 46: ...28 BEFORE USE Getting Started and Exiting 161221687130 承認済 Uncontrolled copy when printed ...
Страница 226: ...208 TEST METHOD Visual Acuity Test by ETDRS chart 161221687130 承認済 Uncontrolled copy when printed ...
Страница 274: ...256 MAINTENANCE List of Replacement Parts 161221687130 承認済 Uncontrolled copy when printed ...
Страница 302: ...284 EMC ELECTROMAGNETIC COMPATIBILITY 161221687130 承認済 Uncontrolled copy when printed ...