TEST METHOD
Binocular Visual Function Test
Horizontal phoria correction
1
Present the horizontal coincidence chart.
Press
.
The polarizing filters are placed in the measuring windows. (The red and green filters are
placed for the SC-2000, SC-1600.)
The base in/out prism powers can be entered.
2
Ask the patient, “Can you see a square with a dot in the center? Are the top and bottom
frames aligned? Or, is one to the left or right of the other?”
When they are not aligned, correct them as follows:
Appearance of
chart
Phoria
Correction
<1> The top frame
is shifted to the left.
Exophoria
Press
. Turn the dial counterclockwise until the
top and bottom frames are aligned.
<2> The top frame
is shifted to the
right.
Esophoria
Press
. Turn the dial clockwise until the top and
bottom frames are aligned.
• The rotary prism lenses are removed from the measuring windows when the already
selected button is pressed again in prism mode.
ex.When
is pressed in the base in/out prism, the rotary prism lenses are removed.
• Use
or
instead of the dial for fine adjustment.
• When the prism power is added, the prism power can be added in 0.5 D increments by
adding the prism power for single-eye compared to both eyes.
Dial 1.0
0.5
,
0.2
0.1
161221687130 承認済
Uncontrolled copy when printed.
Содержание RT-5100
Страница 12: ...X 161221687130 承認済 Uncontrolled copy when printed ...
Страница 18: ... XVI 161221687130 承認済 Uncontrolled copy when printed ...
Страница 46: ...28 BEFORE USE Getting Started and Exiting 161221687130 承認済 Uncontrolled copy when printed ...
Страница 226: ...208 TEST METHOD Visual Acuity Test by ETDRS chart 161221687130 承認済 Uncontrolled copy when printed ...
Страница 274: ...256 MAINTENANCE List of Replacement Parts 161221687130 承認済 Uncontrolled copy when printed ...
Страница 302: ...284 EMC ELECTROMAGNETIC COMPATIBILITY 161221687130 承認済 Uncontrolled copy when printed ...