21 Point Exams (only for Plus Package)
21 Point Exams (only for Plus Package)
5
5.2.3
Gradient AC/A
Setting contents:
Use chart
Near point vertical line chart
Set the near point chart II H (VA1.0) or G (VA0.63) manually.
Auxiliary lens
(R/L)
6
BU/open (rotary prism)
Set power
Subj (N) + Prism value of #13B H.Phoria (N) + (S+1.00)
[Procedure example]
1
After measuring #13B H.Phoria(Near
point), press [Gradient AC/A].
The value becomes S+1.00D added power
automatically and #13B H.Phoria (Near
point) value is set to the rotary prism only
for the first time.
2
Set the near point chart II H (VA1.0) or G
(VA0.63) manually at the 40cm position.
3
Measure the Gradient AC/A with the same
procedures as Steps 2 to 4 in “#13B H.Pho-
ria(N).”
Gradient AC/A
Gradient AC/A is calculated according to
the measured Horizontal Phoria value and
displayed on the bottom center of the win-
dow.