188
TEST METHOD
Binocular Visual Function Test
Stereo test (triangle test)
Use: To detect exophoria or esophoria
Chart: Stereo triangle
Applicable model: SC Pola series G+
Auxiliary lens: Right eye 135º, left eye 45º, polarizing filters
Ideal appearance:
Right eye
Left eye
Binocular ideal
The upper and lower
triangles appear
closer than the central
circle.
1
Present the stereo test (triangle test) chart.
Press
.
The polarizing filters are set in the measuring windows (right eye 135º, left eye 45º).
2
Confirm that the upper and lower triangles appear to the patient to be closer.
The stereoparallax is 10’ 30” (when the refraction distance is 5 m and the PD value is 60
mm, the upper and lower triangles appear to be approximately 101 cm closer to the patient).
The test result OK or NG can be entered by the function button.
3
Press
again.
The direction of the polarizing filters changes (right eye 45º, left eye 135º).
4
Confirm that the upper and lower triangles appear to the patient to be farther away.
They appear to be approximately 171 cm away from the patient (when PD is 60 mm).
5
Repeat Steps 1 to 4 to alternate the triangle appearance.
Only pop-out is slow.
Exophoria tendency
Only recession is slow.
Esophoria tendency
Both pop-out and recession are slow.
Vertical phoria tendency
161221687130 承認済
Uncontrolled copy when printed.
Содержание RT-5100
Страница 12: ...X 161221687130 承認済 Uncontrolled copy when printed ...
Страница 18: ... XVI 161221687130 承認済 Uncontrolled copy when printed ...
Страница 46: ...28 BEFORE USE Getting Started and Exiting 161221687130 承認済 Uncontrolled copy when printed ...
Страница 226: ...208 TEST METHOD Visual Acuity Test by ETDRS chart 161221687130 承認済 Uncontrolled copy when printed ...
Страница 274: ...256 MAINTENANCE List of Replacement Parts 161221687130 承認済 Uncontrolled copy when printed ...
Страница 302: ...284 EMC ELECTROMAGNETIC COMPATIBILITY 161221687130 承認済 Uncontrolled copy when printed ...