66
OPERATING PROCEDURE
Standard Program Refraction
2.5.3
Program #7 (only for Plus Package)
Program #7 is a program to determine the best possible visual acuity with the most plus power based
on the objective measurement data.
Pressing
during the program shows sample instructions to the patient. (The instruction may
not appear depending on the test. At this time, the normal help screen appears.)
1
Enter the AR data.
1) Measure the patient’s eye with an auto refractometer and print the result.
2) Press
.
Confirm that the data No. is the same as the
print data No. of the auto refractometer.
See "2.2.1 From an auto refractometer"
(page 30).
3) Press
.
The AR data appears in the subjective and
AR data fields.
2
Select Program #7.
When the program is not #7, the program can be switched with
+
.
Program #7 can also be displayed by displaying the 21 Point Exams and pressing [#7
Subj(F)] also starts Program #7.
In this case, Step 3 is not necessary.
161221687130 承認済
Uncontrolled copy when printed.