38
OPERATING PROCEDURE
Entering Data
When calling up data from the OPD-Scan III connected with a cable
When entering day data only, see "2.2.1 From an auto refractometer" (page 30).
* For detailed procedure of transferring data from the OPD-Scan III to the RT-5100, refer to the OPD-Scan III
Operator's Manual.
• Set the “AR port” parameter as follows.
When performing patient view simulation with RT data on the OPD-Scan III (V1.05 or later), set
“HDR”. For other cases, set “WF(OPD)”.
• On the Communication tab in the OPD-Scan III Settings screen, set “Connection Device” to
“RT-5100 Support” then check “Send HD Exam Data”.
• On the Settings screen in the Measurement tab of the OPD-Scan III, check “RS232C
Connection” in the Option field.
• Perform Steps 1 and 2 with the OPD-Scan III, and Steps 3 and later with the RT-5100.
1
Measure the patient’s eye with the OPD-Scan III in OPD/CT mode.
For details, refer to the OPD-Scan III Operator’s Manual.
2
When the measured data is obtained, press the Verified and Save buttons of the OPD-
Scan III.
The measured data is printed and automatically stored in the memory of the RT-5100 relay
box.
The number at the top of the printout is called a data No. The number is necessary to call up
the stored data.
• The maximum data count is 9999. However, only the last 150 measurements are retained.
Whenever the data count exceeds 150 measurements, the prior data up to the start of those 150
measurements are deleted.
3
Perform the subjective refinement as with the procedures of Steps 4 and later in “
When calling up data from the ARK-10000 connected with a cable (page 36)”.
When the “AR port” parameter is set to “HDR”, the patient ID of the OPD-Scan III (V1.05 or
later) is also imported.
161221687130 承認済
Uncontrolled copy when printed.
Содержание RT-5100
Страница 12: ...X 161221687130 承認済 Uncontrolled copy when printed ...
Страница 18: ... XVI 161221687130 承認済 Uncontrolled copy when printed ...
Страница 46: ...28 BEFORE USE Getting Started and Exiting 161221687130 承認済 Uncontrolled copy when printed ...
Страница 226: ...208 TEST METHOD Visual Acuity Test by ETDRS chart 161221687130 承認済 Uncontrolled copy when printed ...
Страница 274: ...256 MAINTENANCE List of Replacement Parts 161221687130 承認済 Uncontrolled copy when printed ...
Страница 302: ...284 EMC ELECTROMAGNETIC COMPATIBILITY 161221687130 承認済 Uncontrolled copy when printed ...