205
TEST METHOD
Positive Relative Accommodation (PRA) Test
4
4.12 Positive Relative Accommodation (PRA) Test
Use: To measure the positive relative accommodation when both eyes are converged at a specified
working distance
Chart: Near letter chart with a horizontal line isolation at 40 cm away from the patient’s eye (near point
chart 7)
Confirm that the “SPH Far
Near” parameter is set to SPH + ADD.
For the Plus Package, see "5.2.7 #20 Positive relative accommodation (PRA)" (page 219).
1
Press
while holding
down.
[NPC], [NPA], [NRA], and [PRA] appear.
When the NRA test has been completed just before, skip this step.
2
Press [PRA].
Both measuring windows are opened.
*30
The PRA window is displayed.
3
Gradually turn the dial clockwise and press
[Blur] when the chart appears blur.
The [Blur] indication is highlighted. It
means that the value has been stored.
4
Gradually turn the dial counterclockwise
and press [Recov] when the chart appears
clearly again.
The [Recov] indication is highlighted. It means that the value has been stored.
Pressing [Blur] or [Recov] again clears the stored data.
*30.
Occluding either measuring window with
or
measures the near point of accommodation for either eye.
161221687130 承認済
Uncontrolled copy when printed.
Содержание RT-5100
Страница 12: ...X 161221687130 承認済 Uncontrolled copy when printed ...
Страница 18: ... XVI 161221687130 承認済 Uncontrolled copy when printed ...
Страница 46: ...28 BEFORE USE Getting Started and Exiting 161221687130 承認済 Uncontrolled copy when printed ...
Страница 226: ...208 TEST METHOD Visual Acuity Test by ETDRS chart 161221687130 承認済 Uncontrolled copy when printed ...
Страница 274: ...256 MAINTENANCE List of Replacement Parts 161221687130 承認済 Uncontrolled copy when printed ...
Страница 302: ...284 EMC ELECTROMAGNETIC COMPATIBILITY 161221687130 承認済 Uncontrolled copy when printed ...