• When the patient cannot see the horizontal and vertical lines equally, change the addition
power so that the horizontal lines appear sharper than the vertical ones.
• To continuously perform visual acuity test with addition powers, follow "4.1.3 Visual Acuity
Test with Addition Powers" (page 153).
60
OPERATING PROCEDURE
Standard Program Refraction
19
Check the near visual acuity.
1) Press
.
The cross cylinder lenses are removed.
Set the near point chart II Letters chart at a
position of 40 cm manually.
2) Measure the visual acuity chart for right
eye, left eye, and both eyes.
Press
,
, or
to enter the
VA value by turning the dial.
The VA value cannot be entered with the
mask buttons such as
or
.
20
Select whether the Night measurement test is performed or not.
1) Press
.
The "Night vision test?" message appears.
2) Select whether the Night vision test is
performed or not.
Perform Night vision test
ψ
Select [Yes].
Do not perform Night vision test
ψ
Select [No]. Program
is finished.
Then, Night measurement starts.
The chart becomes the brightness for Night.
• Darken the room for Night measurement. It is recommended to take time to adjust the
patient's eye to the darkness.
21
Check the VA when the chart is darkened
with the full corrected power lens in Day
mode inserted. Perform the subjective
refinement based on the day data and print
the measured result.
1) Press
The isolated 20/20 line is presented.
2) Check the best possible visual acuity
with
or
.
161221687130 承認済
Uncontrolled copy when printed.
Содержание RT-5100
Страница 12: ...X 161221687130 承認済 Uncontrolled copy when printed ...
Страница 18: ... XVI 161221687130 承認済 Uncontrolled copy when printed ...
Страница 46: ...28 BEFORE USE Getting Started and Exiting 161221687130 承認済 Uncontrolled copy when printed ...
Страница 226: ...208 TEST METHOD Visual Acuity Test by ETDRS chart 161221687130 承認済 Uncontrolled copy when printed ...
Страница 274: ...256 MAINTENANCE List of Replacement Parts 161221687130 承認済 Uncontrolled copy when printed ...
Страница 302: ...284 EMC ELECTROMAGNETIC COMPATIBILITY 161221687130 承認済 Uncontrolled copy when printed ...