153
TEST METHOD
Visual Acuity Test
4
4.1.3
Visual Acuity Test with Addition Powers
The VA can be measured with addition powers when the addition powers are entered.
1
Measure the addition powers.
See Steps 23 to 25 in Page 51.
2
Pull down the near point rod and set the near point chart.
Standard: Set either of the near point chart 1 (standard), 5: Letters chart, or 4: Landolt’s
rings.
Plus Package: Set the near point chart II A: Letters chart or E: Landolt’s rings.
3
Press
.
Check the near VA (see Step 26 in Page
51).
4
Press
again.
The VA test with the addition powers starts
and the VA display changes to A-VA.
When the addition powers are not entered,
the test does not start.
Measure the near visual acuity.
5
Test the right, left, and binocular VA.
Press
,
or
and turn the dial
to enter each VA.
The VA cannot be entered with the mask
keys such as
or
.
When LM, AR, Subj, or Final mode is switched during ADD power VA measurement mode,
the mode switches with the VA measurement mode in the ADD power. However, when the
add power is not entered in the specified mode, the mode leaves VA measurement mode in
the ADD power.
161221687130 承認済
Uncontrolled copy when printed.
Содержание RT-5100
Страница 12: ...X 161221687130 承認済 Uncontrolled copy when printed ...
Страница 18: ... XVI 161221687130 承認済 Uncontrolled copy when printed ...
Страница 46: ...28 BEFORE USE Getting Started and Exiting 161221687130 承認済 Uncontrolled copy when printed ...
Страница 226: ...208 TEST METHOD Visual Acuity Test by ETDRS chart 161221687130 承認済 Uncontrolled copy when printed ...
Страница 274: ...256 MAINTENANCE List of Replacement Parts 161221687130 承認済 Uncontrolled copy when printed ...
Страница 302: ...284 EMC ELECTROMAGNETIC COMPATIBILITY 161221687130 承認済 Uncontrolled copy when printed ...