142
ADVANCED FEATURES
Parameter Settings
Auto lamp off: Off, 5 min, 15 min, 30 min
Factory setting: 15 min
Selection of whether or not to use the power saving function. The setting time is also selectable.
The screen backlight and chart presenting device lamp go out in the power saving mode. Press any
key to restore from the mode.
Near lamp link: Yes, Other than grid, No
Factory setting: Yes
Yes
The near point lamp lights up automatically in the ADD mode and near mode.
Other than grid
The near point lamp lights up automatically in the ADD mode and near mode but
the near point lamp goes out when selecting the cross grid chart for near point.
No
The near point lamp does not light up automatically. When lighting up, it is necessary to
press the left side switch of the control box to light up the Near point lamp.
Beep sound: High, Low, Off
Factory setting: High
Setting for the beep sound tone sounded by pressing any key.
Chart link: Yes, No (Other than SC series)
Factory setting: Yes
Selection of whether or not to use the chart presenting device other than NIDEK.
Select No to use the chart presenting device other than NIDEK.
Chart link: Cable, Wireless 1 to 8, No (SC series)
Sets the communication method with the SC series.
Cable
Connects the RT-5100 with the SC series by cable.
Wireless 1 to 8
Communicates with the SC series by an optional infrared communication unit.
The number from 1 to 8 indicates the communication channel. Set the same channel in the
Remote channel parameter of the SC series.
Memory function: Yes, Memory box, No
Factory setting: Yes
Yes
Stores the last 150 data by pressing the print button of the auto refractometer or lensmeter
connected to the RT-5100 with the cable directory.
Memory
Calls up AR or LM data from the optional memory box.
No
Calls up AR or LM data only once without pressing the print button from the auto
refractometer or lensmeter connected to the RT-5100 with the cable directory.
AR port: AR, ARK, RKT, WF(OPD), HDR
Factory setting: AR
Setting for the device to be connected to the AR connector in the relay box.
When connecting the ARK-10000 (V2.11 or later) or OPD-Scan III to AR connector, set the parameter
to “WF(OPD)”.
When performing patient view simulation with RT data on the OPD-Scan III (V1.05 or later), set “HDR”.
LM port: LM, ARK, RKT, AR, WF(OPD)
Factory setting: LM
Setting for the device to be connected to the LM connector in the relay box.
When connecting the ARK-10000 (V2.11 or later) or OPD-Scan III to LM connector, set the parameter
to “WF(OPD)”.
161221687130 承認済
Uncontrolled copy when printed.
Содержание RT-5100
Страница 12: ...X 161221687130 承認済 Uncontrolled copy when printed ...
Страница 18: ... XVI 161221687130 承認済 Uncontrolled copy when printed ...
Страница 46: ...28 BEFORE USE Getting Started and Exiting 161221687130 承認済 Uncontrolled copy when printed ...
Страница 226: ...208 TEST METHOD Visual Acuity Test by ETDRS chart 161221687130 承認済 Uncontrolled copy when printed ...
Страница 274: ...256 MAINTENANCE List of Replacement Parts 161221687130 承認済 Uncontrolled copy when printed ...
Страница 302: ...284 EMC ELECTROMAGNETIC COMPATIBILITY 161221687130 承認済 Uncontrolled copy when printed ...